Disinfection and decontamination:  

• 

Exterior casing and mechanical surfaces must be wiped with a clean cloth, dampened with a disinfectant

• 

Soft plastic parts and rubber surfaces can be cleaned by gently wiping a clean cloth, dampened with a 
disinfectant. Discoloration can occur if alcohol is used

• 

The front lens of eyepieces and objectives are sensitive to chemicals. We recommend not to use aggressive 
disinfectants but to use lens paper or a soft fiber-free tissue, damped in cleaning solution. Cotton swabs may 
also be used. We recommend you use personal eyepieces without eyeshades in order to minimize risk 

• 

Never immerse or dip the eyepiece or objective into a disinfectant liquid! This will damage the component

• 

Never use abrasive compounds or cleaners that may damage and scratch optical coatings 

• 

Properly clean and disinfect all possible contaminated surfaces of the microscope or contaminated accessories 
before storing for future use. Disinfection procedures must be effective and appropriate

• 

Leave the disinfectant on the surface for the required exposure time, as specified by the manufacturer. If the 
disinfectant evaporates before the full exposure time, reapply disinfectant on the surface

• 

Against bacteria, use a 70% aqueous solution of isopropanol (isopropyl alcohol) and apply for at least 30 
seconds. Against viruses, we recommend to refer to specific alcohol or non-alcohol based disinfection products 
for laboratories 

Before returning a microscope for repair or maintenance through a Cole-Parmer dealer, an RMA (return authorization 
form) together with a decontamination statement must be filled in!  This document must be shipped together with the 
microscope at all times.
 

Handle with care

• 

This product is a high quality optical instrument. Delicate handling is required

• 

Avoid subjecting it to sudden shocks and impacts

• 

Impacts, even small ones, can affect the precision of the instrument

Handling the LED 

Note:

 Always disconnect the power cord from your microscope before handling the LED bulb and power unit and 

allow the system to cool down approximately 35 minutes to avoid burns

• 

Never touch the LED with your bare hands

• 

Dirt or fingerprints will reduce the life span and can result in uneven illumination, lowering the optical 
performance

• 

Use only original MSU-100 Series replacement LEDs

• 

The use of other products may cause malfunctions and will void warranty

• 

During use of the microscope the power unit will get hot; never touch it while in operation and allow the system 
to cool down approximately 35 minutes to avoid burns

Dirt on the lenses

• 

Dirt on or inside the optical components, such as eyepieces, lenses, etc., affects the image quality of your system 
negatively

• 

Always try to prevent your microscope from getting dirty by using the dust cover, prevent leaving fingerprints 
on the lenses and clean the outer surface of the lens regularly

• 

Cleaning optical components is a delicate matter. Please, read the cleaning instructions further on in this 
manual

Summary of Contents for 78903-90

Page 1: ...Installation and Operating Instructions MSU 100 Series Compound Microscope Model 78903 90...

Page 2: ...omponents of the microscope____________________________________________________ 7 Preparing the microscope for use __________________________________________________ 8 Assembling Steps _______________...

Page 3: ...a cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before use a...

Page 4: ...based disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination state...

Page 5: ...be between 5 to 40 C and humidity should be within 20 to 80 RH Although the system is anti mold treated installing this product in a hot humid location may still result in the formation of mold or con...

Page 6: ...and the slide The Numeric Aperture N A of the objective is an indication for the resolving power of the objective The total magnification can be calculated by multiplying the magnification of the eye...

Page 7: ...for 3 or 4 objectives B Safety device J Objectives C Tension adjustment K Object stages mechanical stage or clips D On Off switch not visible L Height adjustment lever condenser E Light intensity con...

Page 8: ...r head please pull out the power cord immediately and dry the microscope Assembling Steps Mounting the objectives 1 Rotate the coarse focusing knob to lower the stage to its lowest position 2 Install...

Page 9: ...is lost or damaged choose one with the same specifications Replacement of the Power Cord If the power cord requires replacement ensure that it is suitably rated If in doubt consult your local dealer o...

Page 10: ...ng with a S100x objective you need to lock the slide protection screw D The slide protection screw protects the slide by limiting the travel range of the mechanical stage This way the objectives will...

Page 11: ...h the object stage an Abbe condenser N A 1 25 is mounted The condenser can be adjusted in height by moving the lever beneath the mechanical stage N By adjusting the condenser you can focus the light o...

Page 12: ...er Clean the front lens of the S100x objective 11 Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a drop of isopropanol 12 Clean the slide after us...

Page 13: ...ust can be removed with high pressured air There will never be dust in the objectives if the objectives are not removed from the revolving nosepiece Caution Cleaning cloths containing plastic fibers c...

Page 14: ...as bulbs batteries fuses cords optical components or any add on accessories such as mechanical specimen holders that are not built into the microscope stage as an integral part of the original manufa...

Page 15: ...______________________________________________________ ___________________________________________________________________________________________________ _____________________________________________...

Page 16: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 100_Series_Compound Microscope_Manual...

Reviews: