Use of the S100x oil-immersion objective 

Some Cole-Parmer MSU-100 Series microscopes are equipped with an S100x N.A. 1.25 oil immersion objective. Please 
follow the instructions below on how to use this objective: 

1. 

Remove the dust protection cap from the revolving nosepiece to mount the S100x objective (the objective can 
be premounted)

2. 

Focus the image with the S40x objective

3. 

Turn the revolving nosepiece so the S100x objective almost reaches the click-stop

4. 

Put a small drop of immersion oil on the center of the slide

5. 

Now turn the S100x objective so that you feel the click stop

6. 

The front lens is in contact with the immersion oil

7. 

Look through the eyepiece and focus the image with the fine adjustment knobs

8. 

The distance between the lens of the objective and the slide is very small! 

9. 

In case there are small bubbles visible, turn the S100x objective a couple of times from left to right so that the 
front of the objective moves in the oil and the bubbles will disappear

10. 

After using the S100x objective, loosen the slide protection handle and turn the table with the course 
adjustment knobs downwards until the front lens doesn’t touch the oil any longer. Clean the front lens of the 
S100x objective

11. 

Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a drop of 
isopropanol.

12. 

Clean the slide after use as well

Caution

• 

Never put a drop of xylol or alcohol directly on the lens of the objective. It could enter the objective and dissolve 
the glue that holds the lenses!

• 

Avoid oil contact with any of the other objectives!

Safety device

To prevent damage to the objective lens, or breaking the slide, all types are equipped with a pre-fixed safety device.
It is recommended to use slides of 1.0 – 1.2 mm thickness in combination with cover glasses of 0.13 mm or 0.17 mm 
thickness

Illumination

The LED illumination of the MSU-100 Series is equipped with rechargeable batteries. The length of use after charging is 
about 48 hours. The full charging time is about 10 hours. At first use the batteries will have to be fully charged. Connect 
the external power supply unit to the power inlet socket. For all models with batteries it is highly recommended to use 
the microscope with power supply disconnected to keep the batteries in good working condition and recharge after 
batteries run empty

The illumination has the following specifications:

• 

1W NeoLED  

:  1W, 300mA equipped with Fresnel lens

• 

External Charger  : Primary 100-240 VAC, 50/60 Hz, 0.3 A

• 

Batteries 

: 3 NiMh, AA type, 1.2 Volt  1600 - 2000 mA.

Summary of Contents for 78903-90

Page 1: ...Installation and Operating Instructions MSU 100 Series Compound Microscope Model 78903 90...

Page 2: ...omponents of the microscope____________________________________________________ 7 Preparing the microscope for use __________________________________________________ 8 Assembling Steps _______________...

Page 3: ...a cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before use a...

Page 4: ...based disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination state...

Page 5: ...be between 5 to 40 C and humidity should be within 20 to 80 RH Although the system is anti mold treated installing this product in a hot humid location may still result in the formation of mold or con...

Page 6: ...and the slide The Numeric Aperture N A of the objective is an indication for the resolving power of the objective The total magnification can be calculated by multiplying the magnification of the eye...

Page 7: ...for 3 or 4 objectives B Safety device J Objectives C Tension adjustment K Object stages mechanical stage or clips D On Off switch not visible L Height adjustment lever condenser E Light intensity con...

Page 8: ...r head please pull out the power cord immediately and dry the microscope Assembling Steps Mounting the objectives 1 Rotate the coarse focusing knob to lower the stage to its lowest position 2 Install...

Page 9: ...is lost or damaged choose one with the same specifications Replacement of the Power Cord If the power cord requires replacement ensure that it is suitably rated If in doubt consult your local dealer o...

Page 10: ...ng with a S100x objective you need to lock the slide protection screw D The slide protection screw protects the slide by limiting the travel range of the mechanical stage This way the objectives will...

Page 11: ...h the object stage an Abbe condenser N A 1 25 is mounted The condenser can be adjusted in height by moving the lever beneath the mechanical stage N By adjusting the condenser you can focus the light o...

Page 12: ...er Clean the front lens of the S100x objective 11 Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a drop of isopropanol 12 Clean the slide after us...

Page 13: ...ust can be removed with high pressured air There will never be dust in the objectives if the objectives are not removed from the revolving nosepiece Caution Cleaning cloths containing plastic fibers c...

Page 14: ...as bulbs batteries fuses cords optical components or any add on accessories such as mechanical specimen holders that are not built into the microscope stage as an integral part of the original manufa...

Page 15: ...______________________________________________________ ___________________________________________________________________________________________________ _____________________________________________...

Page 16: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 100_Series_Compound Microscope_Manual...

Reviews: