Technical Manual
Page 36
CODEL
OPS.109
Issue : B
Rev. : 1
Date : 23/10/13
Doc. i/d : 0109/6
Ref. : 090028
5 CALIBRATE
Hi cal 20.0m/sec
Enter
Set the high span calibration target value, 1 digit at a time, using the
and
keys and
Enter
.
5 CALIBRATE
Set # cycles = 50
Set the number of cycles used for calibration, 1 digit at a time, using the
and
keys and
Enter
.
Enter
Calibrate
– manually activates a calibration
5 CALIBRATE
Calibrate
Enter
5 Lo flow 025
Cal in progress
End when count = 0
5 Hi flow 023
Cal in progress
End when count = 0
Config. Cal
– sets parameters for a remote/timed activated calibration
Config Cal
Enter
Initiate
Enter
cal timer
Enter
interval
next cal
cal alarm
logic
output data
Initiate
This enables selection of the time of calibration. Calibration can be triggered by configuring the plant status logic
input to a calibration trigger. Additionally calibration can be triggered at period intervals and pre-set times,
controlled from the on-board clock.