Technical Manual
Page 32
CODEL
OPS.109
Issue : B
Rev. : 1
Date : 23/10/13
Doc. i/d : 0109/6
Ref. : 090028
7.9.5.1. Source
The alarm can be configured to trigger from any one of the four averaged values (sec, mins, hours or days).
5 ALARM Hi
Source 30s
7.9.5.2. Units
The units in which the alarm is operated can be set to m/sec or m
3
/sec.
5 ALARM Hi
Units m/sec
7.9.5.3. Level
The level (threshold) at which the alarm is triggered is adjustable in the
Units
previously selected.
5 ALARM Hi
Level 20m/sec
Select the level, 1 digit at a time, using
and
and set by pressing
ENTER
.
7.9.6. Alarm Lo
The operation of the volt-free contact alarm output (terminals 15, 16 & 17), is configured from this mode.
7.9.6.1. Source
The alarm can be configured to trigger from any one of the four averaged values (sec, mins, hours or days).
5 ALARM Lo
Source 30s
7.9.6.2. Units
The units in which the alarm is operated can be set to m/sec or m
3
/sec.
5 ALARM Lo
Units m/sec
7.9.6.3. Level
The level (threshold) at which the alarm is triggered is adjustable in the
Units
previously selected.
5 ALARM Lo
Level 0.5m/sec
Select the level, 1 digit at a time, using
and
and set by pressing
ENTER
.