3

Calibration 

The output of the Differential Voltage sensor is linear with respect to the input 
voltage. To collect data you can:  
1.

 

Use the calibration supplied by the sensor EEPROM memory.  

2.

 

Use the calibration supplied in the standard sensor library of the Coach program.  

3.

 

Calibrate the sensor using known voltages (a standard, simple 2-point calibration 
can be done). User calibration can be performed in the Coach software. 

 
The name of the Differential Voltage sensor in the sensor library of the Coach 
program is 

Voltage sensor (differential) (0210i) (CMA) (-10 ..10 V)

.

  

 

 
 

 
 
 
 
 
 
 
Figure 2.

  

Default calibration graph of the 
Differential Voltage Sensor (used in the 
Coach sensor library and in the sensor 
memory).  
V

in

 (V)= 1.616 * V

out 

(V) -  0.0081 

where V

in

 = V

+

 - V

-

 

 
Default linear calibration coefficients:  
a=1.616;  b=-0.0081 