4
Technical data
Differential input voltage range
Common-mode input voltage range
± 10 V (voltage between input terminals)
± 50 V (voltage related to ground)
Output voltage range
± 7V
Gain 0.619
x
Resolution using 12 bits A/D converter 7.9 mV
Calibration function
V
in
(V) = 1.616 * V
out
(V) – 0.0081
where V
in
= V
+
- V
-
Input impedance to ground
each terminal 1 M
Input offset voltage error
typical ± 8 mV
Common-mode input voltage error
typical 0.03 mV/V (0 – 500 Hz)
Non-linearity
< 0.001 %
Slew rate
10 V/
s (maximum output voltage variation)
Bandwidth (small signal)
1.2 MHz (-3dB)
Supply voltage
Supply current
5 V DC
typical 19 mA
Maximum differential input voltage
Maximum common-mode input voltage
± 50 V (max. voltage between input terminals)
± 50 V (max. voltage related to ground)
Sensor information for
Auto-ID and calibration
256 byte serial EEPROM
Connection
Right-hand BT (British Telecom) connector
Warranty:
The 0210i Differential Voltage sensor is warranted to be free from defects in
materials and workmanship for a period of 12 months from the date of purchase
provided that it has been used under normal laboratory conditions. This warranty
does not apply if the sensor has been damaged by accident or misuse.
Note:
This product is to be used for educational purposes only. It is not appropriate
for industrial, medical, research, or commercial applications.
Rev. 1/11/12