background image

 

 

 

4

Technical data 

Differential input voltage range 
Common-mode input voltage range 

± 10 V (voltage between input terminals) 
± 50 V (voltage related to ground) 

Output voltage range 

± 7V 

Gain 0.619 

Resolution using 12 bits A/D converter  7.9 mV 

Calibration function  

V

in

 (V) = 1.616 * V

out

 (V) – 0.0081  

where V

in

 = V

+

 - V

-

 

Input impedance to ground 

each terminal 1 M

  

Input offset voltage error 

typical  ± 8 mV  

Common-mode input voltage error 

typical 0.03 mV/V (0 – 500 Hz)                              

Non-linearity 

< 0.001 % 

Slew rate  

10 V/

s (maximum output voltage variation) 

Bandwidth (small signal) 

1.2 MHz (-3dB) 

Supply voltage 
Supply current  

5 V DC 
typical 19 mA 

Maximum differential input voltage 
Maximum common-mode input voltage 

± 50 V (max. voltage between input terminals) 
± 50 V (max. voltage related to ground) 

Sensor information for  
Auto-ID and calibration 

 
256 byte serial EEPROM 

Connection 

Right-hand BT (British Telecom) connector 

 
 
 
Warranty:  

The 0210i Differential Voltage sensor is warranted to be free from defects in 
materials and workmanship for a period of 12 months from the date of purchase 
provided that it has been used under normal laboratory conditions. This warranty 
does not apply if the sensor has been damaged by accident or misuse.  
 

 

Note:

 

This product is to be used for educational purposes only. It is not appropriate 

for industrial, medical, research, or commercial applications.    

 

 

 

 

 

Rev. 1/11/12

 

Reviews: