3

Calibration 

The output of the Current sensor is linear with respect to the input current. To collect 
data you can:  
1.

 

Use the calibration supplied by the sensor EEPROM memory.  

2.

 

Use the calibration supplied in the standard sensor library of the Coach program. 
The name of the Current sensor in the sensor library of Coach is  
Current sensor (0222i) (CMA) (-500 .. 500 mA). 

3.

 

Calibrate the sensor using known currents (a standard, simple 2-point calibration 
can be done). User calibration can be performed in the Coach software. 

 
 

 
 
 
 
 
 
 
 
 
 

Figure 2.

  

Default calibration graph of the 
Current sensor (used in the standard 
Coach sensor library and in the sensor 
memory).  
I (mA) = 78.125 * V

out

 (V) – 0.47

 

 
Default linear calibration coefficients:  
a=78.125;     b=-0.47 
 