Description 

The Current sensor 0222i is designed for measuring small currents in AC and DC 
circuits in the range between -500 and  +500 mA. It has two banana (4-mm) plugs 
for easy connection. 
The sensor contains a sensing element and a signal-conditioning amplifier. The 
sensing element is a 0.4 

 resistor (0.3 W) connected between the red and black 

terminals. As the current passes through the resistor, a small potential difference can 
be measured across this resistor. This potential difference goes through a signal 
amplifier and the output of the sensor is adjusted to the range of 

7.5V, which can 

be measured by an interface. The sensor is protected by a multifuse (resistance of 0.9 

). The time to trip the multifuse to a high-resistant state is 0.1 sec. at 5 A. 

 

The sensor should be connected in series to a circuit element. Currents in either 
direction can be measured. The current is indicated as positive when it flows from 
the red terminal to the black terminal.  

 

The Current sensor is equipped with a BT plug and can be connected to the 
following CMA interfaces: €Lab

CoachLab II/II

+

 and 

U

LAB. Furthermore the 

sensor can be used with Texas Instruments CBL

, CBL2

 and Vernier LabPro.  

Sensor specifications  

The Current sensor has a memory chip (EEPROM) with information about the 
sensor. Through a simple protocol (I

2

C) the sensor transfers its data: name, quantity, 

unit and calibration to the interface

1

Examples of experiments 

The Current sensor is specially designed for measurements of very low and accurate 
currents. It can be used in various experiments such as:  

 

characteristics of a light bulb and a diode, 

 

measurements of internal resonance and EMF, 

 

measurements in series and parallel electrical circuits. 

 
Together with a Differential Voltage Sensor it can be used to explore the 
relationship between the current and the voltage in electrical circuits - the Ohm’s 
Law. 

                                                

 

 

1

 This is valid for the following interfaces: CMA €Lab

BT inputs of CoachLab II/II

+

 and 

U

LAB, 

TI CBL

and CBL2

, and Vernier LabPro.  