Page 34 - FL-20A Instruction Manual
Model FL-20A Cable Fault Locator
Troubleshooting
• An error displays: “The safety switches were let go.”
The two safety switches were released during the test while the Caged
Operation setting was OFF. Press the CANCEL button to return to the
main menu and begin the test again. Hold in the safety switches for the
duration of the test.
• During a shorts test, an error displays: “Inadequate arc rate.”
The voltage may not too low to cause an arc to short from the test con-
ductor to ground. Increase the voltage.
There may be no fault in the product. Run a Hi-Pot test to confirm the
presence of a fault.
• During a shorts test, an error displays: “Accuracy: Low.”
When a test takes a very long time and the report screen says your accu-
racy is low, you may have a multiple fault condition. The FL-20A will find
the first fault so it can be cut out and the cable retested to locate a second
short. This setting is not recommended for low loop resistance products,
because it can give false positives.
• An error displays: “Cannot Find Open.”
If you run the test without attaching the red and blue probes, the error
message, “Cannot Find Open” will be displayed.
•
FL-20A cannot find faults.
In a situation where there are multiple direct or high voltage shorts, you
may not locate any of them.
• The ground probe is connected, but it is not detected.
Insure that both contacts of the ground probe make a good connection
to the ground wire.
• The external interlock is not detected in Caged Operation.
Check the connections from the door or latch on your cage to the inter-
lock pins on the back panel of the FL-20A.
• Test accuracy is rated Low.
Insure that both contacts of each probe are making a good connection.
Insure that the ground probe is connected to all wires that are not being
tested.
The test voltage may be too low.
• The X3F lights are not illuminating when expected.
Check the connections between FL-20A and X3F.
Check the continuity of the connection of each of the testing relay con-
tacts and the connection of each of the HV ON relay contacts. If there
is no connection when expected, replace fuses F1 and F2 on PC board
no. 91496.
• The FL-20A will not power up.
Check the fuse contained in the ON/OFF switch. (See “Maintenance.”)