Clinton 84100 User Manual Download Page 8

METAL

Metal frame and stirrups (if applicable) should be cleaned according to facility antimicrobial standards and dried 
with a clean soft cloth. Never use abrasive cleaners or scrubbing implements on painted surfaces.

UPHOLSTERY 

Clinton uses only premium vinyl for upholstery, however, care should be taken when cleaning to ensure that the 
upholstery or color is not affected by the cleaning agent. Use only mild cleaning agents on upholstery. Clinton 
recommends using a 1 to 1 mix of ammonia and water or a 1 to 4 mix of bleach and water. Rinse thoroughly 
with clean water and pat surface dry. This is not a guarantee and does not relieve the user from responsibility of 
the proper and safe use of all cleaning agents. Test on inconspicuous area prior to using any cleaning agent on 
upholstery's visible surface. Certain cleaning agents can be harmful to the surface.

WARNING:

 Always clean surfaces that come in contact with pathogens or infectious materials  

according to standard procedures of your particular healthcare or therapy facility. Although Clinton 
recommends a cleaning solution it is not a guarantee that it will meet with your facility’s standards 
or state, federal and local standards for treatment and cleaning of areas and equipment that may be 

exposed to pathogens and infectious materials.

PERIODIC INSPECTION 

• Inspect all electrical cords and connections for damage or wear.
• Check that all fasteners are in place and securely fastened.
• Make certain that the hi-lo function of the table is in proper working condition.
• Check to ensure that the table has remained level. 

BACKREST CYLINDER REPLACEMENT 

(Models 84200 & 84430 only)

Remove cotter pin. Then remove  
the slide-out pin. Position tab 
located on the adjustment bar 
against cylinder pin. 

Align holes in cylinder bracket with holes in table bracket by 
adjusting the angle until the holes line up. Install slide-out pin 
through holes and replace cotter pin. 

Finished view of cylinder assembly.

The backrest cylinder is replaceable in the field by following these simple directions below.

Cylinder Pin

 Tab

PREVENTIVE MAINTENANCE & PARTS REPLACEMENT

CLEANING

8

Summary of Contents for 84100

Page 1: ...R I C O P E N B A S E P O W E R TA B L E S C o m b i n e L o a d a n d F o r c e C a p a c i t y 5 5 0 l b s 2 4 9 k g Model Numbers 84100 84200 84430 Limited Warranty ON FRAME ELECTRICAL COMPONENTS Limited Warranty CLI MA 005 REV A ...

Page 2: ...h a comfortable 2 padded surface covered in tough woven back vinyl upholstery The pivot points are self lubricating and the smooth vertical adjustment is powered by a dependable 120v motor Models 84200 and 84430 are equipped with a pneumatic backrest that can be accessed from either side The foot controller on all models features an up button and a down button that controls the operation of the ta...

Page 3: ...ng a problem Pay particular attention to Cautions and Warnings throughout this manual Potentially hazardous and harmful situations to users patients and property can occur if these cautions and warnings are ignored PARTS SAFETY Premium woven back upholstery Plywood Top Seamless rounded corners HEAD END Adjustable Backrest Actuator Lever 1 on each side models 80200 84430 only Top Frame Center Leg C...

Page 4: ...o 90 Operation Temperature 50º F to 100º F Humidity 30 to 75 WARNING When performing any procedure that requires use of electrical instruments or equipment the patient must be insulated fron the metal portions of the table with non conductive material Failure to do so mat result in electrical shock or injury to the patient or attendant WARNING Always use 110v 60hz alternating current Failure to us...

Page 5: ...e the upholstery during the unpacking process CAUTION Clinton Open Base Power Tables are heavy Clinton Industries recommends four reasonably strong individuals using proper lifting techniques be used to lift or move the table Not using the recommended number of movers may result in injury to the persons attempting to lift or move the equipment CAUTION Always lift the table by the metal frame If ta...

Page 6: ...e foot controller with slight even pressure and the table will slowly go down to minimum level Slight even pressure on the up button will slowly raise the table to maximum height CLEAN BEFORE USING All Models It is important to thoroughly clean tables according to facility standards before it is put into service See cleaning suggestions on page 8 of this manual ADJUSTABLE BACKREST Models 84200 844...

Page 7: ... the drop section will lock into position wherever the lifting motion is stopped NEVER LUBRICATE THE LIFT MECHANISM Drop Section in the semi raised position Drop Section in the down position OPTIONAL LOCKING CASTERS All Models Locking casters are an option on all models The locking mechanism can be engaged by using your foot to depressing the tab located at the rear of each caster This action lock...

Page 8: ...ccording to standard procedures of your particular healthcare or therapy facility Although Clinton recommends a cleaning solution it is not a guarantee that it will meet with your facility s standards or state federal and local standards for treatment and cleaning of areas and equipment that may be exposed to pathogens and infectious materials PERIODIC INSPECTION Inspect all electrical cords and c...

Page 9: ...ions and warranties made by any person or entity other than Clinton Industries main office is excluded from this warranty Implied warranties including those of merchantability and fitness for a particular purpose are limited to five years from purchase and all implied warranties are expressly excluded to the extent permitted by law and are otherwise limited to the terms of express warranties as se...

Page 10: ...en nut A and tighten nut B If headrest does not move at all loosen nut B and tighten nut A After proper function has been achieved tighten both nuts Slide Block 4 total Base and Top Frame Lubricate using Molykote G 0052Fg grease top and bottom contact points Base Top Frame End Plug Back Rest Back Rest Actuator Arm Back Rest Cylinder Foot Lift Actuator Arm Motor Center Axle Lubricate using Molykote...

Page 11: ... not covered in this guide or questions about your Clinton Power Table please call Clinton Industries at 800 441 9131 REMOVAL OF THE MOTOR FOR SERVICE The motor assembly is located at the foot end of the table on the base Next detach the motor assembly If possible raise the table to its highest position Raising the table will make it easier to remove the motor mount bolts Using help lower the tabl...

Reviews: