IMPORTANT ELECTRICAL WARNINGS

GENERAL SPECIFICATIONS

 MOTOR SPECIFICATIONS

Models: 84100 & 84200

Table Weight .............................................. 415 lbs.
Total Load Capacity (including force) .......... 550 lbs.
Overall Length ................................................... 72"
Overall Width .............................. 34" & 40" Models
Height Range ...........................................18" to 35"
Backrest 

(Model 84200)

Length ............................................................... 38"
Width ................................................................ 34"
Storage & Transportation
Temperature .................................... 20º F to 150º F
Humidity ............................................-10% to 90%
Operation
Temperature .................................... 50º F to 100º F
Humidity ............................................. 30% to 75%

Models: 84430

Table Weight .............................................. 415 lbs.
Total Load Capacity (including force) .......... 550 lbs.
Overall Length ................................................... 76"
Overall Width .............................. 34" & 40" Models
Height Range ...........................................18" to 35"
Backrest 
Length ............................................................... 38"
Width .....................34" & 40" depending on model
Drop Section
Length ............................................................... 13"
Width .....................34" & 40" depending on model
Storage & Transportation
Temperature .................................... 20º F to 150º F
Humidity ............................................-10% to 90%
Operation
Temperature .................................... 50º F to 100º F
Humidity ............................................. 30% to 75%

WARNING:

 When performing any procedure that requires use of electrical instruments or equipment, 

the patient must be insulated fron the metal portions of the table with non conductive material. Failure 
to do so mat result in electrical shock or injury to the patient or attendant. 

WARNING:

 Always use 110v, 60hz alternating current. Failure to use the proper electrical connection 

may result in electric shock to patient and attendant and will result in damage to the equipment

WARNING:

 Never use these tables in oxygen intense or other explosive environment. Failure to do 

so could result in flash fire or explosion which could result in serious injury or death.

4

Clinton Industries Power Treatment Tables use motors and power supplies that meet UL 60601-1 and 
CAN/CSA C22.2 601.1

.  

 
120VAC    50-60hz    8000N push   1 min. of constant use out of every 9 minutes

Summary of Contents for 84100

Page 1: ...R I C O P E N B A S E P O W E R TA B L E S C o m b i n e L o a d a n d F o r c e C a p a c i t y 5 5 0 l b s 2 4 9 k g Model Numbers 84100 84200 84430 Limited Warranty ON FRAME ELECTRICAL COMPONENTS Limited Warranty CLI MA 005 REV A ...

Page 2: ...h a comfortable 2 padded surface covered in tough woven back vinyl upholstery The pivot points are self lubricating and the smooth vertical adjustment is powered by a dependable 120v motor Models 84200 and 84430 are equipped with a pneumatic backrest that can be accessed from either side The foot controller on all models features an up button and a down button that controls the operation of the ta...

Page 3: ...ng a problem Pay particular attention to Cautions and Warnings throughout this manual Potentially hazardous and harmful situations to users patients and property can occur if these cautions and warnings are ignored PARTS SAFETY Premium woven back upholstery Plywood Top Seamless rounded corners HEAD END Adjustable Backrest Actuator Lever 1 on each side models 80200 84430 only Top Frame Center Leg C...

Page 4: ...o 90 Operation Temperature 50º F to 100º F Humidity 30 to 75 WARNING When performing any procedure that requires use of electrical instruments or equipment the patient must be insulated fron the metal portions of the table with non conductive material Failure to do so mat result in electrical shock or injury to the patient or attendant WARNING Always use 110v 60hz alternating current Failure to us...

Page 5: ...e the upholstery during the unpacking process CAUTION Clinton Open Base Power Tables are heavy Clinton Industries recommends four reasonably strong individuals using proper lifting techniques be used to lift or move the table Not using the recommended number of movers may result in injury to the persons attempting to lift or move the equipment CAUTION Always lift the table by the metal frame If ta...

Page 6: ...e foot controller with slight even pressure and the table will slowly go down to minimum level Slight even pressure on the up button will slowly raise the table to maximum height CLEAN BEFORE USING All Models It is important to thoroughly clean tables according to facility standards before it is put into service See cleaning suggestions on page 8 of this manual ADJUSTABLE BACKREST Models 84200 844...

Page 7: ... the drop section will lock into position wherever the lifting motion is stopped NEVER LUBRICATE THE LIFT MECHANISM Drop Section in the semi raised position Drop Section in the down position OPTIONAL LOCKING CASTERS All Models Locking casters are an option on all models The locking mechanism can be engaged by using your foot to depressing the tab located at the rear of each caster This action lock...

Page 8: ...ccording to standard procedures of your particular healthcare or therapy facility Although Clinton recommends a cleaning solution it is not a guarantee that it will meet with your facility s standards or state federal and local standards for treatment and cleaning of areas and equipment that may be exposed to pathogens and infectious materials PERIODIC INSPECTION Inspect all electrical cords and c...

Page 9: ...ions and warranties made by any person or entity other than Clinton Industries main office is excluded from this warranty Implied warranties including those of merchantability and fitness for a particular purpose are limited to five years from purchase and all implied warranties are expressly excluded to the extent permitted by law and are otherwise limited to the terms of express warranties as se...

Page 10: ...en nut A and tighten nut B If headrest does not move at all loosen nut B and tighten nut A After proper function has been achieved tighten both nuts Slide Block 4 total Base and Top Frame Lubricate using Molykote G 0052Fg grease top and bottom contact points Base Top Frame End Plug Back Rest Back Rest Actuator Arm Back Rest Cylinder Foot Lift Actuator Arm Motor Center Axle Lubricate using Molykote...

Page 11: ... not covered in this guide or questions about your Clinton Power Table please call Clinton Industries at 800 441 9131 REMOVAL OF THE MOTOR FOR SERVICE The motor assembly is located at the foot end of the table on the base Next detach the motor assembly If possible raise the table to its highest position Raising the table will make it easier to remove the motor mount bolts Using help lower the tabl...

Reviews: