CI-90A User’s Manual
Page 31
5.2 Calibration service
The CI-90A, like any other electronic measurement equipment, should be calibrated at least
annually. The purpose of the calibration is to establish that the unit is in tolerance, and that
sample volume is therefore accurate.
The accuracy and consistency of your CI-90A can be degraded by unqualified and marginally
experienced calibration services. To insure that the flow rate of the instrument is maintained to
factory specifications, you may ship your unit to either our Eastern or Western service facility,
or arrange for on-site service. For locations outside the U.S., consult your local distributor.
CLiMET’s calibration service follows ANSI Z-540 and ISO 10012, providing both
as found
and
as left
measurements, as well as
in-tolerance
or
out-of tolerance
findings.
Our Service Center can be contacted by phone at 909-793-2788, by fax at 909-793-1738, or
by email at
When shipping units, use the original shipping carton, a well padded shipping case, or a box
that allows at least two inches of bubble wrap extending from the greatest protrusion on
all
sides
of the instrument. Loose packing material, commonly referred to as
peanuts,
provides
psychological comfort, but very little real protection.
Summary of Contents for CI-90A
Page 2: ...CI 90A User s Manual Page ii ...
Page 6: ...CI 90A User s Manual Page vi ...
Page 12: ...CI 90A User s Manual Page 6 ...
Page 13: ...CI 90A User s Manual Page 7 ...
Page 14: ...CI 90A User s Manual Page 8 ...
Page 16: ...CI 90A User s Manual Page 10 ...
Page 30: ...CI 90A User s Manual Page 24 ...
Page 34: ...CI 90A User s Manual Page 28 ...
Page 42: ...CI 90A User s Manual Page 36 ...
Page 44: ...CI 90A User s Manual Page 38 ...