Message
Purpose o
f Message
Ty
pe
Action Required
Rel. Blood Volume
Low
The current blood volum
e
percentag
e o
r hematocrit
has
drop
ped bel
o
w
the BV Ale
rt
Level set on the “Crit-Line
”
screen. The
UF pump ha
s
been turn
ed
off.
Alert
Assess
the patient for any
changes in physiologic state.
The alert can be cleared b
y setting the
BV Alert Level
button on
the “Crit-
Line” screen
to OFF.
To turn the BV Alert lev
el
OFF, selec
t t
he
BV Alert Level
button,
enter “0” (zer
o) on the keyboar
d, and p
ress the
CONFIRM
key
.
After clearing the alert, th
e UF pump will automatically turn back on
to the UF Rate set on the
“Home” scre
en.
Note
: At the
direction of a physician, a
new BV Alert Level can b
e
selected.
To change th
e BV Alert Level, select the
BV Alert Level
button,
enter the des
ired valu
e using the keyb
oard a
nd pre
ss
the
CONFIRM
key
.
Verify
Failed
The v
erification test
shows
that the CLiC device is not
read
y for use during
treatment.
Advisory
message
Verify the CL
iC device:
Place the CLiC device on
the CLiC
Verification Fil
ter, which is
attached to its USB
cable.
Wait up to one minute for the messag
e to clear. If the message
clears, the C
LiC device is verified.
If
the messa
ge is not cle
ared:
Select the “Test & Options”
screen-b
utton then select
the
Verify
Crit-Line
button. This will start
the verification process manually.
Wait up to one minute for the messag
e to clear.
If
the messa
ge is still not
cleared, the
CLiC device cannot be used for the
dialysis treatment.
Disconnect the CLiC de
vi
ce cable from the USB p
ort.
Press
the
R
ESET
key to
disable the CLiC device.
Contact a qualified
service technician.