2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
23
Ending a CLiC device monitored treatment
When the treatment is complete and the Tx Clock is paused, pinch the
CLiC device to spread the sensor elements apart and gently remove the
device from the Crit-Line Blood Chamber. Store the CLiC device by
clipping it to its verification filter, which is attached to its USB cable.
Discard the disposable Crit-Line Blood Chamber with the rest of the
bloodlines.
Printing a treatment report
When RTD is zero and the Tx Clock is paused, a
button will appear
on the “Crit-Line” screen in place of the
Markers
button.
To clear the data on the “Crit-Line” screen without printing, press the
New
Tx
key and then press the
CONFIRM
key.
See the “CLiC Spare Parts” section on page 32 for information on kits
necessary to connect and use a printer.
The image on the facing page is an example of a CLiC device monitored
treatment printout; below is an explanation:
Item
Description
S/N
The CLiC device’s four digit serial number (printed on the
device’s verification filter).
Patient ID
The first 6 digits of the “Test & Options” screen Patient ID
number. This is the unique identifier for this report.
Note
: If the Patient ID is fewer than 6 digits long, or if there is
no Patient ID, this field will display six ones (“111111”).
Hct (start)
The initial hematocrit (Hct) value is recorded along with the
first data point when graphing begins.
Hct (max)
The highest level of hematocrit during treatment.
Hct Limit
The hematocrit limit (indirectly set based on the BV Alert
Level on the “Crit-Line” screen).
~Hgb (start)
The estimated initial hemoglobin (Hb) value recorded along
with the first data point when graphing began.
Sat (min)
The minimum oxygen saturation value over all data points.
Station ID
Not applicable.
Date
The date of the treatment.
Start Time
The time the dialysis treatment began.
Stop Time
The time the dialysis treatment ended.
Calibration Date The date the CLiC device was factory calibrated.
Last Verification
Date
The last date the CLiC device was verified for accuracy in
Dialysis Mode.
HCT Graph
Hematocrit graph
%BV Graph
Blood Volume graph
SAT% Graph
Oxygen Saturation graph