2008T Hemodialysis Machine with CLiC Device User’s Guide
P/N 490206 Rev E
17
The O
2
Sat (Oxygen Saturation) graph
Oxygen saturation is the default display on the “Crit-Line” screen’s bottom
graph.
Note
: If blood pressure readings are displayed instead of the
measured oxygen saturation, touch or select the Blood Pressure
graph on the screen. The message “
Press Confirm to switch to
O2 graph. Press Escape to return to BP graph.
” is
displayed. Press the
CONFIRM
key to confirm the selection.
To switch back to the Blood Pressure graph, select the graph
again and press the
CONFIRM
key to confirm the selection.
The O
2
Sat
(
Oxygen Saturation) graph is a line graph drawn from the
previous oxygen saturation value. The O
2
Alert Level is drawn as a green
dotted line across the Oxygen Saturation graph.
The following is displayed to the right of the Oxygen Saturation graph:
Feature
Function
Current Oxygen Saturation value.
Minimum Oxygen Saturation – The minimum oxygen
saturation value over all data points.
O
2
Alert Level
button – This button sets the oxygen saturation
alert level. When the current O
2
Sat value drops below the O
2
Alert Level, the alert message “Oxygen Saturation Low” is
displayed in the Status Box. The alert can be cleared by
lowering the alert level or by setting the O
2
Alert Level to OFF.
To set the O
2
Alert Level, select the
O
2
Alert Level
button and
enter the desired value using the
or
(up or down) keys or
the numbers on the keyboard and press the
CONFIRM
key.
To turn the O
2
Alert level OFF, enter “0” (zero) on the
keyboard and press the
CONFIRM
key.
The O
2
Alert Level defaults to 89 and can be set from 45 to 95.
Setting the O
2
Alert Level to any value 44 or lower turns the O
2
Alert Level OFF.