26
P/N 49020
6 Rev E
2008T
Hem
odialysi
s Mach
ine with CLiC Device
User’
s Guid
e
Message
Purpose o
f Message
Ty
pe
Action Required
Check cable or press
‘Y’ to di
sable Crit-Line
The 2008T h
emodialysis
machine has not received
data from the CLiC de
vice.
The Status Box displays the
warnin
g message “Crit-Li
ne:
No Comm”
Advisory
message
Check the CLiC device’s USB cable:
Make sure the cable is
securely conne
cted to the 2
008T
hemodial
ysis machine’s U
SB port.
Wait up to one minute for the messag
e to clear.
If
the Status
Box warni
ng
message “Cr
it-
Line: No C
omm” is not
cleared, the
CLiC de
vice cannot be used fo
r the dia
lysis
treatment.
Disconnect the CLiC de
vi
ce cable from the USB p
ort.
Press the ‘Y’ key on the Keybo
ard
to di
sable the CLiC device.
Contact a qualified
service technician.
Crit-Line dis
abled.
Must v
erify
Crit-Line.
The CLiC de
vice must be
verified befo
re it is availab
le
for use durin
g treatment.
Advisory
message
If
this me
ssa
ge is displayed
durin
g the
treatment
:
The CLiC de
vice is not
monito
ring the
patient. The CLiC de
vice
cannot be used until the next treatment.
If
the messa
ge is displayed
before th
e treatme
nt starts
:
Place the CLiC device on
the verifi
cation filter, which is attached to
its USB cable.
Wait up to one minute for the messag
e to clear. If the message
clears, the C
LiC device is verified.
If verification
fails, select
the “Test & Options” screen-button an
d
select the
Verify
Crit-Line
button to initiate the verification process
manually.
If
the messa
ge is still not
cleared,
the CLiC device cannot be used
for the dialysis treat
ment.
Disconnect the CLiC de
vi
ce cable from
the USB port and contact
a qua
lified service technician.
C
rit-
Lin
e ne
ed
s
ve
rification
The CLiC de
vice has not
been
verified withi
n the past 30
days. The Hct/BV graph on
the “Crit-Line” screen will
be
disabled.
Advisory
message
Verify the CL
iC device:
Place the CLiC device on
the CLiC
Verification Fil
ter, which is
attached to its USB
cable.
Wait up to one minute for the messag
e to clear. If the message
clears, the C
LiC device is verified.
If
the messa
ge is not cle
ared:
Select the “Test & Options”
screen-b
utton then select
the
Verify
Crit-Line
button to start th
e verification process ma
nually.
Wait up to one minute for the messag
e to clear.
If
the messa
ge is still not
cleared, the
CLiC device cannot be used for the
dialysis treatment.
Disconnect the CLiC de
vi
ce cable from the USB p
ort.
Contact a qualified
service technician.