Troubleshooting
1.
The Pump Fails To Operate
•
If attempting to run from mains power supply:
♦
Check that the mains power supply lead is firmly connected
and that the power is on at the mains power supply outlet.
•
If attempting to run from the internal battery, check the
Battery Status LED on the Control Panel. If it is on (Red) refer
to the Charging the Internal Battery section.
•
Ensure that the Vacuum Gauge on the Control Panel indicates
zero. If not, release vacuum and switch pump off and on.
2.
Pump Running But No Vacuum
•
Check that all fittings are connected tightly.
•
Turn the vacuum controller knob on the control panel and
watch the indicator on the vacuum gauge.
•
Ensure that the silicone medical suction tubing is in good
condition and not old and cracked.
•
Check the type of handpiece in use as some handpieces
require finger occlusion or rigger release to allow suction.
•
Check bacteria/hydrophobic shut off valve filter.
If the pump fails to work and you are unsure why, contact your
service department or return unit to your distributor.
Troubleshooting