Cleaning
Cleaning
After Each Operation
Empty Collection Jar
1.
Switch off pump.
2.
Disconnect tubing from jar “PATIENT “ port.
3.
Disconnect tubing from jar “VACUUM” port.
4.
Carefully remove jar from pump and dispose of aspirated
fluid according to local protocols for potentially
biohazardous waste.
Clean
Housing
Wipe clean with a damp soapy cloth. Do not immerse or allow
liquid to enter the housing. Do not use abrasive cleaning agents.
Suction Tubing
Suction Tubing should be sterilised with sterilants determined by
local protocols and compatible with silicone tubing. Suction
tubing may be autoclaved to a maximum of 121°C for 15 minutes.
At higher temperatures the suction tubing will discolour and lose
shape.
Collection Jars
Place jars upright or up-side-down in autoclave -
not on their
sides
. Autoclave to a maximum of 121°C for 15 minutes. Do not
use phenolic solutions as disinfecting agents in polycarbonate
jars.
Bung (Lid)
Unlock bayonet on float cage by rotating cage
clockwise
.
Disconnect float cage and float from bung (lid) and separate seal
from bung (lid). Place all items upright or up-side-down in
autoclave -
not on their sides
. Autoclave all items to a maximum
of 121°C for 15 minutes.