Important Information:
Before using the Clarity BG1000 Blood Glucose Monitoring System, please read this manual
carefully.
The Clarity BG1000 Blood Glucose Monitoring System is for testing outside of the body (in-vitro
diagnostic use)
.
This product is not for neonatal use or severely dehydrated individuals or patients.
Use this system only for blood glucose test.
Make sure to use your test strip immediately after retrieving it from the vial,
be sure to keep the test strip vial closed tightly at all times
.
For the most accurate results, make sure your hands are clean and dry before removing the
test strips from the vial. Do not touch test strip slit.
Do not use test strips that have expired, using expired test strips may cause inaccurate test
results.
Do not reuse test strips. This test strip is for single use only.
Before the test you must make sure that the code number displayed on the meter matches the code
number on the test strip vial.
Be sure to check the glucose measurement unit on the meter before the test.
- This meter is preset to mg/dl.
- mg/dl is the unit of measurement used in the US.
Make sure the test strip vial is closed tightly and stored at 39
.
2-86°F
.
Do not
freeze.
Keep test strips away from direct sunlight and heat.
The meter and test strips should be handled at the same temperature
.
For accurate test results, keep the Clarity BG1000 Blood Glucose Monitoring System
operating temperature at a range between 50-104°F for more than twenty (20) minutes
before testing
.
You may use this system at altitude up to 10,000 feet.
Keep your Clarity BG1000 Blood Glucose Meter, lancing device, and control solutions out of
reach of children.
Clean and disinfect your Clarity BG1000 Blood Glucose Meter before testing each patient.
Dispose of used test strips and lancing device according to local regulations. Inaccurate results may
occur for individuals or patient in shock.
Inaccurate results may occur for individuals experiencing a hyperglycemic- hypersmolar state with
or without ketosis
.
Critically ill patients should not be tested with this blood glucose meter
.
Interferences: Acetaminophen, salicylates, uric acid, ascorbic acid (vitamin C) and other interferent
substances in normal blood or normal therapeutic concentrations do not significantly affect results,
however abnormally high concentrations in blood may cause inaccurate results
.
Patients undergoing oxygen therapy may yield false low results.
Lipemic samples (triglycerides) in excess of 1,500 mg/dl may produce elevated results.
Hematocrit is the percentage of red blood cells in the blood
.
HGT levels of 30-55% do not affect
glucose measurements with this meter
.
This test strip is not for use with arterial, venous, neonatal, serum or plasma samples.
Testing outside of HGT level range of (30-55%) and recommended storage conditions may cause
inaccurate results
.
Testing under out of the specification range may cause inaccurate results.