PROCEDURE MANUAL
____________________________________________________________________________
Product Code: CD-BG1
Rev01.06042018
Page
14
of
20
Understanding Control Solution Test Results:
- The example label shown above on the test strip vial shows the acceptable ranges for Level 2 and
Level 3 Control Solution for the Clarity BG1000 Glucose Meter.
- The result you get should be within this range. Make sure you compare the result to the correct level
of control.
- Each batch of test strips has different acceptable ranges.
Cleaning and Disinfecting:
The meter should be cleaned and disinfected to prevent the transmission of blood borne
pathogens as follows
.
Cleaning is to remove germs
,
dirt
,
and impurities from surfaces or
objects
.
Disinfection is to kill germs on surfaces or objects;
- User should wear disposable gloves when cleaning and disinfecting the meter.
- CaviWipes
®
has been validated for use with this meter to clean and disinfect the outside
of the meter. The towelette is available for purchase through your distributor. Also you can
get this product from large online retail distribution sites such as www.amazon.com ,
www.officedepot.com , etc.
- CaviWipes® disinfectant products have been shown to be safe for use with the Clarity
BG1000 Blood Glucose Monitoring System. Please note, any disinfectant product with the
EPA registration number of 46781-8 may be used on the Clarity BG1000 Blood Glucose
Meter.
- The Clarity BG1000 Blood Glucose Monitoring System has demonstrated that there was no change
in performance or in the external materials of the meter after 11
,
000 cleanings with disinfection cycles
designed to simulate three years of device use.
CAUTION:
Do not use other cleaners or disinfectants because they may damage the meter.
Cleaning Instructions:
Use CaviWipes
®
towelette to wipe the meter exterior of the front, the back and the sides
thoroughly
.
Disinfection Instructions:
The meter must be cleaned and disinfected between patient uses by wiping it with a CaviWipes
®
Towelette in between tests. After cleaning the meter with a CaviWipes
®
Towelette use a second
Caviwipes
®
Towelette to thoroughly wet the pre-cleaned
meter surface
.
Allow it to remain wet for two
minutes at 68°F as specified by the
manufacturer of the disinfectant wipes and allow to air dry
.