Your Safety Tester
6
Specifications
Input Rating
Supply Voltage
.............................................................. 120 Vac RMS
Frequency
.............................................................................. 50/60 Hz
Fuse
..................................................... 20 Amp thermal circuit breaker
Ground Bond Test
Test Voltage
.................................................................6 Vac (Nominal)
Test Current
....................................................................... 7A (Typical)
Fail Level
................................................................................... 500m
Ω
Short to Case Test
Test Voltage
.................................................................6 Vac (Nominal)
Test Current
........................................................... 20
μ
Amps (Typical)
Fail Level
..................................................................................... 150k
Ω
Short to Line Test
Test Voltage
.................................................................6 Vac (Nominal)
Test Current
................................................................. 1 Amp (Typical)
Fail Level
........................................................................................... 3
Ω
Leakage Test
Test Voltage
............................................................................. 120 Vac
Fail Leakage Current
.............................................................. 0.75 mA
Accuracy /-10% at 22
°
C and 40% relative humidity.
This product is IP40 rated.