Nitrogen Generator
NiGen HF - NiGen LCMS
35
the following page will appear
Active errors are those highlighted; once the error is identified, follow the in-
structions in the following paragraph.
To reset the alarm, press the RESET key
9.3.
List of warnings
warning identification
cause
resolution
W-00
: data loss
the memory has lost the set
up values of the parameters
and has restored the default
values
contact service in order to
set the correct values of the
parameters
W-01
: low nitrogenoutlet
pressure
outlet nitrogen pressure
below threshold
reduce outlet nitrogen flow
W-02
: maintenance requi-
red
maintenance counter has
decreased to zero
check maintanance sche-
dule; perform the reque-
sted maintenance operation;
reset related maintenance
counter
W-03
: low air outlet pres-
sure
outlet air pressure below
threshold
(only for NiGen
LCMS 40-1 or NiGen HF-1)
reduce outlet air flow