Nitrogen Generator
NiGen HF - NiGen LCMS
31
8.3.
Maintenance schedule
e
If a correct maintenance schedule is not followed, the performance of the gene-
rator may no longer be ensured and it may lead to permanent damage to the
generator. The following table specifies the frequency of the recommended
maintenance
operations, express in “working hours” of the compressor.
e
Lack of maintenance will void the warranty.
m
maintenance operations must be performed by authorized personnel
8.3.1.
NiGen HF-1 and NiGen LCMS 40-1
Components use versus maintenance plan
Component part
number
Description
2000 h
4000 h
6000
8000
R0539021
Intake air filter cartridge
1
1
1
1
R0539022
Intake air filter cap
1
1
1
1
R062999001
Silicone disk NRV
16X6mm
1
1
R0539023
Motor capacitor
1
R063399
Filter cartridge Syntesi 5
micron
1
1
1
1
R063400
Filter cartridge Syntesi
20 micron
1
1
1
1
R0539024
Dumping feet kit (3 pie-
ces)
1
1
R061084
CMS silencer 3/8
1
R700.M4.7060
Compressed air cooler
1
R0539025
Maintenance compres-
sor kit
R0539027
Maintenance compres-
sor kit
Maintenance kit to be used -->
422.93.0025
422.93.0026