GUS Disinfection Soak Station™
M0036_F
8
G32-S / G32-E
Filter Life
GUS®
has a patented filter that, in normal everyday use, will have a six-month life. Call
CIVCO Medical Solutions at 1-800-445-6741 to purchase additional filters
Filter Replacement
CAUTION Turn the main power OFF and disconnect the power cord from the
electrical outlet.
NOTE:
It is recommended that the filter change take place at the time of replacing
the disinfectant.
1.
Turn the main power switch
OFF
and disconnect the power cord from the electrical
outlet.
2.
Remove the top louver plate
(see figure 2)
located at the top of the machine by
unscrewing the acorn nut with a 5/16 nut driver. Note the G32 has a single top louver
plate.
3.
Remove the old filter. Put the used filter into a plastic bag and dispose of it with other
non-hazardous waste. This does
not
need to be red tagged.
4.
Install new filter and replace louver panel.
Figure 2: Top Louver Plate