GUS Disinfection Soak Station™
M0036_F
12
G32-S / G32-E
APPENDIX B: SITE REQUIREMENTS
Environmental Conditions
•
For indoor use only.
•
Altitude up to 2,000 meters.
•
Temperature range: 41 °F to 104 °F.
•
Maximum relative humidity: 80% for temperatures up to 87.8 °F, decreasing linearly to
50% relative humidity at 104°F.
•
Main supply voltage fluctuations not to /- 10% of the nominal voltage.
•
Insulation over voltage category II.
•
Pollution degree 2.
Supply Ratings
AC 110-120V 0.5A - 2.0A 50/60 Hz
Fuse: 250V 3.0 A 5X20 FSF Panel Mounted
Main Power/Ratings Plate
The main power switch is located on the inside front panel of the system (
see figure 3
). The
switch is illuminated when the system is ON. The power cord plugs into a regular wall outlet.
The fuse is a 250V 3.0 A 5X20 FSF, panel mounted.
The ratings plate and serial number are located on the right side of the system.
Figure 3: Front Panel of the System
Actual Size
Model
Air
Treated
(CFM)
FPM
w/door
closed
low
FPM
w/door
open
high
Voltage
Current
w/door
closed
Current
w/door
open
Dimensions
(outside)
H x W x D
Dimensions
(inside)
H x W x D
G32-E
220
120
50
110V
50/60Hz
40 watts
75 watts
26.5”H
33.25”W
27.5”D
22”H
32”W
22”D
G32-S
220
120
50
110V
50/60Hz
40 watts
75 watts
16.5”H
33.25”W
27.5”D
14”H
32”W
22”D
Suction Hose Fitting
for Disposal System
Disposal System
On/Off Switch
Ground Fault
Circuit Interrupter
Main Power On/Off
Switch