GUS Disinfection Soak Station™
M0036_F
9
G32-S / G32-E
CAUTION The system’s carbon filter must be replaced every six (6)
months.
Prior to changing the filter, neutralize and dispose of any high-level
disinfectant in the soaking tray(s), and remove all equipment from the
system.
Disconnect the system’s power supply before removing the filter.
CAUTION An F321-A filter is required for use with hydrogen peroxide
or peracetic acid based high level disinfectants.
Cleaning the System
WARNING
When cleaning the system, always wear appropriate personal
protective equipment. Potentially infectious material adhering to the
unit could pose an infection control risk.
CAUTION Do not use solvents to clean the acrylic door. Doing so may
damage the doors. Do not wipe surfaces with hard or abrasive
materials. Doing so will scratch the surface
1.
Turn the power switch
OFF
and disconnect the power cord from the wall outlet.
2.
To clean metal surfaces: Use a clean, lint-free cloth moistened with a mild detergent
solution and water. A mild solvent may be used on these surfaces for difficult-to-
clean areas
3.
To clean the HDPE plastic (side panels), use mild detergent solution such as dish
washing liquid and water. Hard to clean spots may be cleaned with a mild solvent if
extra care is taken.
4.
Make sure that the unit is completely dry before use.
Rinse Water
Once removed from the disinfectant solution, the instruments/scopes must be thoroughly
rinsed with water to remove any disinfectant residue. The initial rinsing
must
be conducted in
the GUS
®
system in order to eliminate risk of exposure to residual chemicals left on the
instrument
NOTE:
Refer to your instrument or scope’s manufacturer instruction manual
regarding the use of sterile water.