doseBadge User Manual
Page 53
Appendix 3 Resetting the doseBadge and Reader Unit
Resetting the doseBadge
If, in the event that the doseBadge does not respond correctly to commands
from the Reader Unit and if all of the troubleshooting steps have been
followed, this reset procedure can be followed.
This procedure should only be followed if the doseBadge does not respond to
commands from the RC:112A Reader Unit. Please ensure that all of the troubleshooting
listed above has been followed before carrying out this procedure.
In particular, ensure that the section covering the Shake-to-Wake function has
been read and understood.
Please contact Cirrus Research plc or your local distributor for further details.
The purpose of this procedure to perform a hard reset on a doseBadge. This procedure
should only be followed, if you have been instructed to do so by Cirrus Research, on an
individual unit.
The actions described in this procedure should not erase any data from your doseBadge
unit, but Cirrus Research plc do not take any responsibility for loss of data or any
damage that occurs to your doseBadge from following this procedure.
Please fully read and make sure you understand the whole procedure before starting it.
If you do not feel confident about following the procedure, please do not attempt it!
Warnings
This symbol is used throughout this procedure to indicate that special
attention should be taken and any instructions followed carefully.
1.
With a small, high quality Philips PH00 screwdriver remove the three screws from
the base of the doseBadge.
2.
Place the metal label and three screws to one side.
SCREWS