![Cirrus Research doseBadge CR:112A User Manual Download Page 13](http://html.mh-extra.com/html/cirrus-research/dosebadge-cr-112a/dosebadge-cr-112a_user-manual_2608628013.webp)
CR:112A & CR:112AIS doseBadge User Manual
Page 13
Charging the doseBadges
Please note that this manual refers to the CU:110A Charger Unit only. The
information given does not apply to any previous version of the CU Charger
Units. The CU:110A is identified by a Black Case with the older charger units
having a White Case.
The CU:110A Charger Unit has a maximum charge time of 2½ hours. If the doseBadge
is not fully discharged, the charge time will be shorter than this.
It is essential that this procedure is carefully followed otherwise damage will occur to
the doseBadge battery.
Before using the doseBadges for the first time, ensure that the units are allowed to
charge fully. The doseBadges should be regularly recharged to maintain the
performance of the battery pack (inside the doseBadge)
For further information regarding the charging and maintenance of the doseBadge,
please refer to Page 38 “Maintenance & Care”
Connect the Charger to the mains power supply and switch on the power supply.
Ensure the doseBadge is in Stop Mode and carefully screw it onto the "charging stud"
on a CU Series Charger unit.