21.
Launch a GOSUN scenario in the CU menu. The sensor head is going
to point the sun. Polar and Azimuthal values are displayed in degree on
the CU screen.
After a GOSUN scenario the sun spot may not be centered on the target.
The GPS coordinates, UT time and date being received automatically from the satellite,
the error on the polar axis should be inferior to 1 mm as long as the spirit level is well
centered (If superior refer to part 5.2.2).
If not centered on the azimuthal axis, refine the manual positioning of the robot base
as done in step 14. (The screws being loosened the base can still be moved).