19501 Series Fixture B195000/B195001/B195002 Quick Start Guide
15
cover is closed, the Interlock terminal will be
short-circuited.
The functions of the B195002 rear terminals are shown in the figure
below.
(1)
Interlock
: Please use the enclosed wire to connect to
the Interlock terminal on the high-voltage
module of the 19501 partial discharge tester.
The test starts only when the acrylic cover is
closed making the micro switch short-circuit.
(2)
GND
: Please use the enclosed ground wire to
connect the GND terminal on the high voltage
module
of the 19501 partial discharge tester.
The installation of B195002 DIP test fixture on the 19501 partial discharge
tester is shown in the figure below:
Use the enclosed M3*20mm screws to secure the B195002 test fixture on
the high voltage module of the
19501 partial discharge tester. The
direction and holes position are shown in the figure below.