The high voltage module supporter is used to prevent the module from
vibration during testing. The supporter has magnets for fixation. Follow the
arrows shown in the figure below to fix the supporter.
Note
When the
high voltage module
of the partial discharge tester is
installed with
an
EMI shielding can
and set the INTERLOCK-2 to
ENABLE
, besides short-circuiting the two terminals of INTERLOCK-1
during high voltage output, the INTERLOCK-2 switch must be pressed