Service setup
H Series Service Manual
15
020-000802-03 Rev. 1 (12-2016)
After a few seconds, the ADC calibration status will change from Doing to Done.
Video calibration
To preform the video calibration, complete the steps below.
1. On the IR remote keypad, press Menu > Size & Position > Size Presets > Full Size.
2. Enter Factory Mode, see
Entering factory mode
on page 9.
3. Select ADC Calibration > Help.
After a few seconds, the ADC calibration status will change from Doing to Done.
1
Test equipment
Video generator
(output level: 700 mv)
2
Test signal
720P
3
Test pattern
SMPTE BAR
4
Inspection
items
Check if the ADC calibration status has changed from Doing to Done.