Manufacturer’s guide and declaration – Electromagnetic Immunity
Immunity test
IEC 60601 test level
Level of compliance
Electromagnetic
environment - guide
Electrostatic discharge
(ESD) IEC 61000-4-2
In contact ± 6 kV
In air ± 8 kV
In contact ± 6 kV
In air ± 8 kV
Floors should be wood, con-
crete or ceramic tile. If the
floors are covered with syn-
thetic material, the relative hu-
midity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power lines
± 2 kV for power lines
The quality of the network
voltage must be equivalent
to that of a typical commer-
cial or hospital setting.
Surge
IEC 61000-4-5
± 1 kV between phases
± 2 kV between phases
and earth
± 1 kV between phases
± 2 kV between phases
and earth
Voltage
failures,
short
interruptions and voltage
variations in the power
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycles.
40% U
T
(60% dip in U
T
)
for 5 cycles.
70% U
T
(30% dip in U
T
)
for 25 cycles.
5% U
T
(95% dip in U
T
) for 5 s.
<5% U
T
(>95% dip in U
T
) for
0.5 cycles.
40% U
T
(60% dip in U
T
)
for 5 cycles.
70% U
T
(30% dip in U
T
)
for 25 cycles.
5% U
T
(95% dip in U
T
) for 5 s.
Magnetic field at mains
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Electromagnetic fields at
network frequency must
have levels equivalent to
those of a typical commercial
or hospital setting
NOTE: U
T
is the network AC voltage before applying the test level.
Conducted disturbances,
induced by radio-
frequency fields
IEC 61000-4-6
3 V eff
150 kHz at 80 MHz
V1=3 V eff
150 kHz at 80 MHz
Radiated radio-frequency
IEC 61000-4-3
3 V/m
80 MHz at 2.5 GHz
E1= 3 V/m
80 MHz at 2.5 GHz
(*1) Portable and mobile RF communications devices should not be used near any part of the Naturally ME Electric
Breast Pump Mod. 06836
00009199000000, including the cables. Further information on distances to be kept be-
tween portable and mobile radio communications equipment and the Naturally ME Electric Breast Pump Mod. 06836
00009199000000 may be requested from Artsana S.p.A., using the contact information provided in this manual. It is
nevertheless advisable to keep the appliance at a safe distance (at least 1 m) from mobile phones or RF transmitters in
order to reduce possible interference.
NOTE 1: These guidelines may not cover all situations. Electromagnetic radiation is affected by absorption and reflection
of structures, objects and people.
This product complies with the Directive 2012/19/EU.
The crossed bin symbol on the appliance indicates that the product, at the end of its life, must be disposed of sepa-
rately from domestic waste, either by taking it to a separate waste disposal site for electric and electronic appliances
or by returning it to your dealer when you buy another similar appliance. The user is responsible for taking the ap-
pliance to a special waste disposal site at the end of its life. If the disused appliance is collected correctly as separate
waste, it can be recycled, treated and disposed of ecologically; this avoids a negative impact on both the environment and
health, and contributes towards the recycling of the product’s materials. For further information regarding the waste disposal
services available, contact your local waste disposal agency or the shop where you bought the appliance.
WARRANTY
The product is guaranteed against any conformity defect in normal conditions of use as provided for by the instructions for
use. The warranty shall not therefore apply in the case of damages caused by improper use, wear or accidental events. For
the duration of warranty on conformity defects please refer to the specific provisions of national laws.
Summary of Contents for 00 009199 000 000
Page 2: ...2 ...
Page 3: ...3 A 1 8 9 2 5 3 4 6 7 18 19 ...
Page 4: ...4 B C 10 17 14 12 11 16 15 13 20 24 22 23 21 ...
Page 5: ...5 1 5 3 7 2 6 4 1 2 ...
Page 113: ...113 NOTE ...
Page 114: ...114 NOTE ...
Page 115: ......