Problem
Cause
Solution
The breast pump does not pump or
pumps very little.
Valve not assembled or not properly
assembled.
Assemble the valve correctly.
Cup and motor body not assembly
correctly.
Check that you have performed
the procedure described in the
“PREPARATION” paragraph correctly.
Make sure that the silicone tube is
firmly connected to both the motor
unit and the membrane cover.
Check that the tube is not damaged.
Do not twist the tube during milk
expression.
The cup does not fit well on the breast
Fit the pump body over your breast,
taking care that your nipple is
properly centred. Try with/without
the soft silicone cover.
Suction level not correctly adjusted.
Change the suction level until
you achieve more comfortable
stimulation and suction by pressing
the “+” or “-” buttons.
Run-down and/or insufficiently
charged batteries.
Replace the batteries.
Milk expression is painful.
Suction level too high.
Change the suction level until
you achieve more comfortable
stimulation and suction by pressing
the “-” button.
The breast pump does not switch on
Failure of mains adapter and/or
appliance.
Check that mains power is present.
Check the position of the batteries in
the battery compartment.
Incorrect connections.
Check all the connections.
Run-down batteries.
Replace the batteries.
If none of the above operations solves the problem, contact Artsana S.p.A.
TECHNICAL SPECIFICATIONS
- Ref. 00 009199 000 000 ELECTRIC BREAST PUMP Mod. 06836
- Maximum vacuum level in extraction mode: -0.35 ± 0.03 bar**
(** Values measured with sample cylinder. CI 95%)
- Appliance service life: 500 hours
- Accessory service life: 2 years
- Service life in storage conditions: 4 years
- Noise level: less than 50 dBA at a distance of 1 m
- Type of appliance: intermittent operation; 30 minutes ON - 30 minutes OFF
- Motor body weight: 165 g excluding batteries
- Dimensions: 98x70xh105 mm
- Made in China
Storage, preservation and transportation conditions:
- Ambient temperature: -20 / +60°C
- Relative humidity: 10 to 85%
- Pressure: 700 to 1060 hPa
Operating conditions:
- Operating temperature: +5°C / +40°C