Chemtrac Auto Clean UV254 Owner'S Manual Download Page 15

Chemtrac UV254 Monitor 

– UVM5000 v 2.1 

 

Chemtrac, Inc.  

 

14 

 

5.7 

Electrical Connections 

The 24 VDC power is supplied to the UVM5000 by means 
of a DC power adapter that is plugged into a wall outlet of 
100-240 VAC 50/60 Hz. The two 24 VDC wires from the 
adapter are run through the connector in the bottom of the 
monitor. The black wire is connected to terminal #11 (-) 
and the red wire (+) is connected to terminal #12, as 
shown in Figure 5.14. 

Both 24 VDC power and communication lines feed into the 
monitor via one ¾” electrical conduit fitting on the bottom 
of the monitor and must be connected to the terminal block 
on the main control board.  

 

5.8 

4-20 Ma Connections and Dry C ontacts 

Primary communication is provided by a 4-20mA 
sourcing current loop output. RS-232 is also included to 
allow direct connection to a PC COM port if desired. A 
configurable digital output / dry contact is also 
available. These provide the operator with alarms for 
operator configured UVT/UVA set-points and system 
failure alarms. Dry contact connections are from 
terminals #5 & #6. The 4-20 mA output connections are 
from negative terminal #3 and positive terminal #4. 

Dual Feed Connection

 

– For the dual feed option, a 

second pair of 4-20 mA output connections are from 
negative terminal #1 and positive terminal #2. This 
output signal is for the water stream 2. 

Automatic Chemical Clean Connection

 - Power to 

the optional Automatic chemical clean system is 
provided at connector terminals #9 and #10, which run 
to the left terminal block on the circuit board in the 
chemical clean cabinet. Matching the correct polarity is 
important. The chemical clean logic is through 
connector terminals #7 & #8, which run to the right terminal 
block on the circuit board in the chemical clean cabinet. 
Polarity is not important. 

 

 

Note:

 Terminal numbering from left to right 

Figure 5.14: Power connections 

Table 4: Main Terminals 

Stream 2 4-20mA 

– (dual feed option only)  

Stream 2 4-20mA + (dual feed option only)  

4-20mA 

–  

4-20mA +  

Dry Contact Out 

Dry Contact Out 

Clean Control Out  

Clean Control Out  

24VDC Out 

– (Black) (for use with Auto Clean) 

10 

24VDC Out + (Red) (for use with Auto Clean) 

11 

24VDC In - (Black)  

12 

24VDC In + (Red) 

 

Summary of Contents for Auto Clean UV254

Page 1: ......

Page 2: ...Chemtrac UV254 Monitor UVM5000 v 2 1 Chemtrac Inc 1 Chemtrac Inc 1555 Oakbrook Dr Suite 100 Norcross GA 30093 770 449 6233 chemtrac chemtrac com...

Page 3: ...Connections ___________________________________________________________ 14 5 8 4 20 Ma Connections and Dry Contacts______________________________________________ 14 6 Start up _______________________...

Page 4: ...________________________________ 37 16 Parts _____________________________________________________________________________ 38 17 Cabinet Dimensions and Clearance ______________________________________...

Page 5: ...amount of light able to pass through two different path lengths of test water From these two measurements alone the UV transmittance UVT or UV absorbance UVA of the test water is calculated without co...

Page 6: ...Contact Chemtrac for replacement parts 2 Ensure that all responsible personnel carefully read this manual before installing or servicing this analyzer 3 This analyzer contains a UV lamp which emits li...

Page 7: ...acklit LCD Indicator light for system alarms and warnings ALARMS Operator configurable alarms for high and low set points low lamp output leak detected system fault etc HUMIDITY CONTROL Humidity senso...

Page 8: ...ll support up to 50lbs per component 6 Sufficient space and clearance to access and service the system Refer to drawings in mounting section to ensure appropriate location is selected Minimal distance...

Page 9: ...e the UVM5000 enclosure 10 mm x 20 mm Quartz Flow Cell packed inside the UVM5000 enclosure 1L Calibration Bottle with Flex Tubing 24 VDC Power Adapter and Cable packed inside the UVM5000 enclosure Deh...

Page 10: ...middle region of the flow cell where the light beam passes through Wearing gloves is recommended 2 Insert the flow cell into position under the flow cell header as shown below in Figure 5 4 Align the...

Page 11: ...t until you feel it click into place Figure 5 8 3 Press the top and bottom hold down clamps against the connector to anchor it into position red arrows in Figure 5 8 4 Attach the other end of the ribb...

Page 12: ...5 5 Plumbing Process water from a pressurized pipe flows into the monitor via a push to connect fitting The sample is analyzed by the sensor inside the cabinet and discharged to drain as shown in Figu...

Page 13: ...g from the process to the IN push connector on the monitor 2 Connect another piece of tubing to the OUT port on the monitor and divert the line to the drain IMPORTANT The test water must be allowed to...

Page 14: ...surement This results in a measurement of organics material independent of turbidity and particulate in the water The same results can also be achieved by removing light blocking particulate from the...

Page 15: ...tem failure alarms Dry contact connections are from terminals 5 6 The 4 20 mA output connections are from negative terminal 3 and positive terminal 4 Dual Feed Connection For the dual feed option a se...

Page 16: ...ing through the flow cell and no water droplets are on the inside of the monitor 4 Plug in the power supply Display screen will read Initializing 5 Calibrate your Analyzer Measurements are not assumed...

Page 17: ...tubing at the inlet connection on the monitor 5 Fill the supplied wash bottle with DI RO water and connect the bottle tubing to the inlet connection on the monitor 6 Hold a finger over the small vent...

Page 18: ...Press Enter 3 A default value is displayed in the New UVT field in this example 75 0 4 Press the up arrow to increase the number above 75 0 or the down arrow to decrease the number below 75 0 When yo...

Page 19: ...r scaling Default 100 4mA UVA Press Enter then up down to select 4mA UVA reading for scaling Default 0 20mA UVA Press Enter then up down to select 20mA UVA reading for scaling Default 1 4mA Crltn Pres...

Page 20: ...ating arm Default 1 On REF POSITION Factory settings Steps Press Enter then up down to change motor alignment increasing the adjusts the position clockwise RS 232 DATA LOG Setup Baud Rate 9600 Fixed p...

Page 21: ...tus 0 no alarm 1 alarm Clean Fluid Low Alarm status 0 no alarm 1 alarm System Fault Alarm status 0 no alarm 1 alarm SET ALARMS See Dry Contact 0 Disabled 1 Alarm Humidity Press Enter then up down to s...

Page 22: ...r then up down to select 0 to disable or 1 to enable Default 0 Idle Time Press Enter then up down to select number of minutes to wait while switching between sample lines Default 2 Cycle Time Press En...

Page 23: ...ty The 4 20 mA can be set to produce a user selectable UVT or UVA value for the low 4 mA and a separate value for high 20 mA up to 110 UVT Figure 10 1 below is an example of a 4 20 mA setup for a cust...

Page 24: ...communication protocol for RS232 is ASCII An example of one data stack is shown here based on the use of optional turbidity compensation The order of the data stack in ASCII from left to right is UVT...

Page 25: ...he data stack in ASCII from left to right is UVT 1 UVT 2 Removal Lamp 1 Water 1 VIS 1 Lamp 2 Water 2 VIS 2 Temperature Humidity Status Table 6 itemizes each parameter and its corresponding value Witho...

Page 26: ...Lab Value 4 The relationship between the UVA value and the desired water quality parameter can then be determined using the formula on the right 5 Alternatively the slope can be determined via multip...

Page 27: ...to an hour to properly dry the air inside the monitor Each time the door to the cabinet is opened this drying process starts over so it is recommended to keep the cabinet door closed as much as possi...

Page 28: ...as a starting point upon installation Wastewater 4 24 hour frequency Source Water 1 7 day frequency High Purity 2 4 week frequency Once in operation observe the flow cell for signs of fouling prior to...

Page 29: ...Acid 30 solution Rust Off Hagasen Blue Removes lime calcium rust and color staining Citric Acid 20 solution Removes mineral scaling Sodium Hypochlorite 6 solution Removes oil grease and biofilm Acetic...

Page 30: ...monitor 6 Once the fluid starts to flow remove finger from the vent hole and place bottle on top of the monitor as shown in Figure 14 6 The fluid should continue to flow through the monitor If it stop...

Page 31: ...ecause some water may still be in the cell Slowly lift the blue lever of the quartz flow cell clamp pictured in Figure 14 7 on the right 4 Grip the quartz flow cell with two fingers at the top and car...

Page 32: ...To replace the UV lamp follow these steps 1 Unplug the monitor from the power supply 2 Shut off the water flow 3 Drain the flow cell by disconnecting the tubing from the external push in fittings on...

Page 33: ...14 Check for leaks If a leak occurs immediately shut off the water supply and drain the flow cell and then reinstall it 15 Power the monitor 16 Close the door and watch the Secondary Data entry in th...

Page 34: ...rom time to time Chemtrac recommends replacing the blue ribbon cable every 12 18 months as preemptive maintenance The replacement is a very simple procedure requiring no tools 1 Unclip the end connect...

Page 35: ...circuit board If blown replace with 20mm x 5mm 1 1 0 0 Bad reference position check Check reference microswitch on motor bracket If damaged it may need to be replaced Check set screw attaching rotati...

Page 36: ...e door especially after shipping Lamp High The lamp output is too high and is out of range of the sensor If a new lamp has been installed it may be necessary to adjust the internal gain of the sensor...

Page 37: ...ator setpoint Flow cell is dirty and needs to be cleaned OR a grab sample should be obtained and compared to the sample water to confirm measurement 15 3 Secondary Data The Secondary Data menu gives r...

Page 38: ...numbers are dependent on the absorbance of the sample water stream CONDITION VALUE ISSUE ACTION Lamp values within range Less than 100 Sample is close to the bottom of the measurement range OR flow c...

Page 39: ...0 Calibration Bottle Plastic Calibration Bottle for UV254 analyzers 25155 Standard Power Supply 24 VDC Power adapter and cable for UV254 analyzers 25170 NEMA 4 IP65 Power Supply 1 6A NEMA 4 IP65 rated...

Page 40: ...imensions and Clearance The front side and back views of the cabinet are shown below to the nearest 1 8 inch The back view gives the dimensions relevant for spacing of screws to mount the cabinet Allo...

Page 41: ...lure caused by 1 Improper installation operation or maintenance of equipment 2 Abnormal wear and tear on moving parts caused by some processes 3 Acts of nature i e lightning etc This warranty represen...

Reviews: