Chemtrac UV254 Monitor
– UVM5000 v 2.1
Chemtrac, Inc.
10
5.3.3
Installing the Ribbon Cable
Note:
Ensure not to pinch the ribbon cable when installing as this
can create a weak spot in the cable and reduce the life of the
cable.
1. Position the flex ribbon connector to the rotating arm as
shown in Figure 5.7.
2. Press the connector into the slot until you feel it click into
place (Figure 5.8).
3. Press the top and bottom hold down clamps against the
connector to anchor it into position (red arrows in Figure
5.8).
4. Attach the other end of the ribbon cable into the slot at the
control board and then secure the top and bottom clamps
against the connector (Figure 5.9).
Figure 5.8: Snap receptacle into place; secure clamps
Figure 5.9: Attach cable to control board
Press clamp down
so that it is flush
with connector
Opposing clamp
hidden from view.
Press up.
Figure 5.7: Orient ribbon cable