5 Operation of the NucleoCounter
17
4.
Close the lid gently.
After analysis the lid is lifted and the cassette is removed
from the insertion slit and disposed of according to national
or regional laws or regulations regarding the nature of the
mixture it contains.
Figure 12.
Illustration of how to grip and insert the cassette
into the insertion slit
5.3
Measuring - The Run button
When a cassette has been loaded and inserted into the
instrument, the measurement is activated, by pressing the
Run
button.
5.3.1
Actuator movements
When
Run
is pressed, the actuator moves the piston rod down
through the cylinder of the cassette. Hereby, the sample/dye
mixture is transported to the cassette measurement chamber. The
form and dimensions of the flow channels facilitate effective
mixing of sample and stain. Sensors monitor sample flow. When
the sample has been correctly loaded into the measurement
chamber, the piston is stopped.
Summary of Contents for NucleoCounter
Page 1: ...NucleoCounter User s Guide P N 991 0009 Revision 1 4 Technology that counts...
Page 2: ...i This page was intentionally left blank...
Page 4: ...iii This page was intentionally left blank...
Page 6: ...v This page was intentionally left blank...
Page 8: ...vii This page was intentionally left blank...
Page 10: ...ix This page was intentionally left blank...
Page 13: ...Introduction and intended use xii This page was intentionally left blank...
Page 21: ...1 Inspection and Unpacking of Equipment 4 This page was intentionally left blank...
Page 28: ...Fejl Henvisningskilde ikke fundet 11 This page was intentionally left blank...
Page 42: ...5 Operation of the NucleoCounter 25 This page was intentionally left blank...
Page 49: ...6 Total and viability count 32 This page was intentionally left blank...
Page 53: ...7 Maintenance of NucleoCounter 36 This page was intentionally left blank...
Page 58: ...8 Troubleshooting Error messages 41 This page was intentionally left blank...