background image

Page 10 

 
 

4.3.3  CAN Protocol. 

Two separate protocols are available from the setup – either multiple or single message. For the 
former, the microDAQ is allocated a fixed message for each group of 4 pressure channels, ie for a 
64  channel  scanner,  16  discrete  CAN  message  identifiers  are  required.    Alternatively  for  a  more 
economical use of identifiers within a system, a single message id may be used, with all channels 
sent  over  this  message  sequentially.  Channel  numbers  are  positively  identified  with  a  channel 
identifier byte within the message. 
 
For the multiple message option, data are sent on 8 byte messages, 4 channels per message – the 
pressure  channels  associated  with  a  particular  message  number  being  fixed.  The  2  most 
significant  digits  of  the  message  identifier  are  user  selectable  via  the  setup  program  –  so  for 
example the identifier for channels 1-4 for user selected 0x22n would be 0x220, and the identifiers 
for channels 5-8, 9-12 etc. follow on incrementally from this first identifier as shown in Figure 4.3.  It 
is  the  user’s  responsibility  to  avoid  overlap  of  message  identifiers  in  a  multiple  microDAQ 
installation. 
 
Data are two bytes unsigned, scaled to +/- full scale, ie 0x0000 might represent –5psi, 5psi.  
The data are user selectable as 16 bit big or little ended.  An example of the packing is illustrated in 
Figure  4.3.    Use  of  the  16  bit  little  ended  option  is  recommended  as  it  requires  least  processor 
overhead within the microDAQ unit. 
 
The  single  message  identifier  option is included to reduce the number of message id’s required 
within  a  system.    Data  are  packed  in  7  byte  messages  as  3  channels  per  message  with  one 
channel  identifier  byte.    This  identifier  byte  is  incremented  for  each  message  in  the  sequence, 
starting at 0x00 for channels 1,2 3, then 0x01 for channels 4,5,6 etc.  Odd leftover channels (eg 16 
channels/3 = 5 full messages plus a message with a single channel and two ‘leftovers’) are set to 
0x0000  and  should  be  discarded.    Figure  4.4  shows  the  structure  of  a  message  for  the  single 
message  identifier  protocol.    A  selectable  inter  message  delay  of  between  1ms  and  200ms  is 
available  to  match  message  generation  timing  to  the  user’s  system.    Note  that  it  is  the  user’s 
responsibility to select appropriate delays with respect to channel rate and CAN bus baudrate. 

Summary of Contents for microDAQ

Page 1: ...Folgate Road North Walsham Norfolk NR28 0AJ ENGLAND Tel 01692 500555 Fax 01692 500088 microDAQ Pressure Scanner Acquisition System USER PROGRAMMING GUIDE e mail info chell co uk Visit the Chell websit...

Page 2: ...ified in this manual may impair the user s protection Chell Document No 900167 Issue 1 3 ECO 1270 Date 28th November 2014 Chell s policy of continuously updating and improving products means that this...

Page 3: ...ol 7 4 1 4 Control Via RS232 7 4 1 5 Data Rate 7 4 2 TCP 8 4 2 1 Overview 8 4 2 2 Connection 8 4 2 3 TCP Protocol 8 4 2 4 TCP Data Rate 8 4 2 5 Control Via TCP 9 4 3 CAN 9 4 3 1 Overview 9 4 3 2 CAN B...

Page 4: ...received command may be acknowledged if required The following sets out the essentials of the command protocol the data packet format for all channels in addition to the message identifier arrangemen...

Page 5: ...ASCII The response to the command is to echo back the string Test command rxd ok n where n is the value of the parameter byte This may be checked from a terminal program by typing dC ie byte values 6...

Page 6: ...pressure value is averaged over the time between data delivery cycles Rezero and Rebuild G 71 None Requests a rezero followed by a calibration table rebuild with the calculated zero offsets being app...

Page 7: ...rnal RAM Disable the data delivery for the chosen channel Get Status 63 0 Short 1 With temp 2 Full 3 Pressure reading 4 Temp readings 5 Excitation reading 6 Hall sensor read 7 Firmware ID 8 Unit seria...

Page 8: ...channels A calibration table rebuild is then performed Hardware Trigger T 84 byte 0xab a 0 Disable a 1 Enable b 0 RS232 b 1 TCP b 2 CAN b 3 Internal RAM b 4 Internal RAM stop on full Enables or disabl...

Page 9: ...in ascending order as comma delimited ASCII engineering units ie degrees C The full status data contains the above followed by fields for the setup options of the microDAQ Each field is comma delimite...

Page 10: ...ations bandwidth as well as less processor overhead within the microDAQ it is recommended that engineering unit conversions be applied at the client Protocol Example data format Note 16 bit LE 0x00 0x...

Page 11: ...ing to 10kHz from the default maximum 20kHz due to internal issues regarding string handling The 16 bit little ended protocol is most efficient from the microDAQ s viewpoint the data being calibrated...

Page 12: ...account to avoid losing data Please note that if streaming at high data rates it is essential that a good network infrastructure is used It is recommended that any streaming is performed over a priva...

Page 13: ...ty to avoid overlap of message identifiers in a multiple microDAQ installation Data are two bytes unsigned scaled to full scale ie 0x0000 might represent 5psi 0xffff 5psi The data are user selectable...

Page 14: ...MSB CH18 MSB CH22 MSB CH26 MSB CH30 MSB 2 CH2 LSB CH6 LSB CH10 LSB CH14 LSB CH18 LSB CH22 LSB CH26 LSB CH30 LSB 1 CH1 MSB CH5 MSB CH9 MSB CH13 MSB CH17 MSB CH21 MSB CH25 MSB CH29 MSB 0 Ch1 LSB Ch5 LSB...

Page 15: ...The offset from the base identifier may be selected as being 0x10 0x20 0x30 0x40 or 0x50 For example the base data message identifier of 0x220 might be set up to receive commands over CAN on message...

Page 16: ...on the selected CAN message type single message or multiple message IDs as detailed in section 4 3 3 above In addition to the above an internal RAM dump sends a 9 or 6 byte header which is sent as soo...

Page 17: ...ion of the standard command set and can be transmitted all through the same comms connection These commands change the valves accordingly to perform a zero function or a purge function and also to sim...

Reviews: