USER MANUAL
13
4.12
Factory settings
If zeroing with standard plastic films does not work, it may be necessary to reinitialize the device. First
press briefly
select "Reset", then "Complete Reset" and confirm with
"Complete reset" will flash in the display and indicate together with a beep that the initialization is
completed. To ensure high measurement accuracy, it is recommended to perform a zero point calibration
after initialization. All previously saved values and settings are deleted after initialization and the device is
reset to factory settings.
Note
:
When measuring the coating thickness of alloys such as aluminum or stainless steel alloy, the
instrument must be calibrated as follows:
1. Turn on the coating thickness gauge, initialize it to factory settings, and turn it off.
2. Turn it back on and calibrate it to 6 standard slides (if accuracy was affected, keep the unit away
from any material during the entire calibration process).