background image

User Manual

        

LDM100 

Coating Thickness Gauge

 

Checkpoint Instruments

ul. Klecińska 125

54-413 Wrocław 

Polska

kontakt

@

cp-instruments.pl

www.

cp-instruments.pl

Summary of Contents for LDM100

Page 1: ...User Manual LDM100 Coating Thickness Gauge Checkpoint Instruments ul Klecińska 125 54 413 Wrocław Polska kontakt cp instruments pl www cp instruments pl ...

Page 2: ... measurement 6 4 3 2 Continuous measurement 8 4 4 Measuring range 8 4 5 Battery indicator 8 4 6 Selection of the measuring unit 9 4 7 Language selection 9 4 8 Alarm function 10 4 8 1 Switch on off the alarm 10 4 9 Storedvalues 11 4 10 Accuracy measurement 11 4 11 Calibration 12 4 11 1 Zero point calibration 12 4 11 2 Multi pointcalibration 12 4 12 Factory setting 13 5 Disposal 14 6 Contact 14 ...

Page 3: ...evice for the first time The device may only be used by carefully trained personnel This meter may only be used in the manner described in this manual Use of the meter elsewhere may result in dangerous situations The device must not be exposed to extreme temperatures direct sunlight extreme humidity or moisture The device housing may only be opened by qualified personnel of the CP Instruments The ...

Page 4: ... mil 0 1 µm 0 1 mil 2 2 µm 2 0 1 mil 166 x 68 x 30 mm 180 g with batteries 10 50 C Note Perform a measurement only when the probe has reached the specified ambient temperature Please do not observe the first measurement results if you are not sure whether the sensor has already reached the specified ambient temperature ...

Page 5: ...Height Press the short in measuring mode to the statistics data Max Min Average values displayed by the device This screen will be displayed in the next picture Data Enter Press the short in measuring mode to display the recorded data This screen will be displayed in the next picture Calibration Backlight In measurement mode press and hold to turn on the backlight Then you can or adjust brightenes...

Page 6: ...rument A beep sounds and Model no PCE CT 27FN is displayed on 4 3 1 Single measurement SI mode is the measurement mode of the single measurement Press the probe vertically against the surface to be measured The unit will beep once and display the coating thickness value when the readings have been taken Note Do not pull the probe over the surface to be measured The screen will be shown in the next...

Page 7: ...USER MANUAL 7 standard deviation 20 5 µm 2 2 µm Fe The measured material is a magnetic metal such as steel NFe The measured material is a non magnetic metal such as aluminum ...

Page 8: ...ure 4 4 Measuring range The device measures layer thicknesses between 0 μm and 1250 μm If the thickness is outside this range will appear in the display The screen will be shown in the next picture 4 5 Battery indicator The device is powered by four AAA batteries If the icon is green the battery is fully charged After a period of use the battery icon will become shorter indicating the current capa...

Page 9: ... to the menu item Unit via the main menu for to select between μm and mil the options menu Press the 4 7 Lanugage selection Press the long to get to the main menu Then move to the menu item Language via the main to select between English EN Chinese CN and German DE menu for the options menu Press the ...

Page 10: ... or long for fast scroll to adjust the range 4 8 1 Switch on off the alarm Press the long to get to the main menu Then move to the menu item Alarm via the main menu for the options menu Press to turn the alarm on or off When the alarm is on the unit displays the following picture on the display If the measurement is outside the alarm limits the unit will beep three times to signal this ...

Page 11: ...vice according to the given reference standards The device is supplied with standard plastic films These can be used to test the measurement accuracy of the instrument with the help of supplied calibration blocks They can also be used to protect the sensor on rough or hot surfaces The measured value should be within the accuracy range specified in the operating instructions For example if the accu...

Page 12: ...ickness measurements Nevertheless due to abnormal base materials or extreme environmental conditions additive errors may occur If this is the case you can use the standard plastic films to perform a multipoint calibration 4 11 2 Multipoint calibration Use the instrument to place standard plastic films on a calibration block if the value found is outside the accuracy limits of this manual Lift the ...

Page 13: ...ommended to perform a zero point calibration after initialization All previously saved values and settings are deleted after initialization and the device is reset to factory settings Note When measuring the coating thickness of alloys such as aluminum or stainless steel alloy the instrument must be calibrated as follows 1 Turn on the coating thickness gauge initialize it to factory settings and t...

Page 14: ... Checkpoint Instruments ul Klecińska 125 bud BETA 54 413 Wrocław For the implementation of the ElektroG take back and disposal of WEEE we take back our equipment They are either recycled by us or disposed of by a recycling company according to legal requirements 6 Contact For questions about our product range or the measuring device please contact CP Instruments kontakt cp instruments pl Postal Ch...