4.1.13 Leak's & Power
Figure 4.45: Leak’s & Power measurement menu
Test results / sub-results
P ................... Active power
Itou ............... Touch Leakage current
Idiff ............... Differential Leakage current
S ................... Apparent power
Q .................. Reactive power
PF ................. Power factor
THDu ............ Total harmonic distortion – voltage
THDi ............. Total harmonic distortion – current
Cos Φ
...........
cosinus Φ
I .................... Load current
U ................... Voltage
Test parameters
Duration
Duration [Off, 2 s ... 180 s]
Change status
Change [YES, NO]
YES: The instrument measures leakage current in two sequential steps with delay* in
between. The phase voltage is firstly applied to the right live output of the mains
test socket and secondly to the left live output of the mains test socket.
NO: The phase voltage is applied only to the right live output of the mains test socket.
*Delay time
Delay [0.2 s … 5 s]
Output connections
[Socket L–N, Socket L,N – PE,P]
Test limits
H Limit (P)
H limit [Off, 10 W ... 3.50 kW, Custom ]
L Limit (P)
L limit [Off, 10 W ... 3.50 kW, Custom ]
H Limit (Idiff)
H limit [Off, 0.25 mA ... 15.0 mA, Custom ]
L Limit (Idiff)
L limit [Off, 0.25 mA ... 15.0 mA, Custom ]
H Limit (Itou)
H limit [Off, 0.25 mA ... 15.0 mA ]
L Limit (Itou)
L limit [Off, 0.25 mA ... 15.0 mA ]