Test circuit
Figure 4.40: Measurement of Touch Leakage current
Touch Leakage measurement procedure
Select the
Touch Leakage
function.
Set test parameters / limits.
Connect device under test to mains test socket. Connect test lead to P/S terminal on the instrument and on
device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.41: Examples of Touch Leakage measurement results
4.1.12 Power
Figure 4.42: Power measurement menu