16
HV AC measurement procedure
Select the
HV AC
function.
Set test parameters / limits.
Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.
Connect HV test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.12: Examples of HV AC meaasurement results
Note:
First HV measurement after power on the instrument (if password protection is enabled) or first HV
measurement after enabling or changing password require entering password for enabling HV test.
4.1.4 HV DC
IMPORTANT SAFETY NOTE
Refer to chapter
1.1 Warnings and notes
for more information regarding safe use of the instrument.
Figure 4.13: HV DC test menu
Test results / sub-results
U ................... measured test voltage
I .................... test current
Test parameters
DC test voltage
U test [500 V ... 6000 V in steps of 50 V]
Duration
t end [Off, 1 s ... 120 s]