Test results / sub-results
P ................... Active power
S ................... Apparent power
Q .................. Reactive power
PF ................. Power factor
THDu ............ Total harmonic distortion – voltage
THDi ............. Total harmonic distortion – current
Cos Φ
...........
cosinus Φ
I .................... Load current
U ................... Voltage
Test parameters
Duration
Duration [Off, 2 s ... 180 s]
Output connections
[Socket L–N]
Test limits
H Limit (P)
H limit [Off, 10 W ... 3.50 kW, Custom ]
L Limit (P)
L limit [Off, 10 W ... 3.50 kW, Custom ]
Test circuit
Figure 4.43: Measurement of Power
Power measurement procedure
Select the
Power
function.
Set test parameters / limits.
Connect device under test to mains test socket.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.44: Examples of Power measurement results