33
The unit has two charging modes:
- fast charging mode : battery recovers 90% of its capacity in 3h ;
- service charging mode : this mode is initiated at the end of the fast charging mode, or before
when the battery is very low, or during a measurement if the adapter is connected.
Note : The charging mode (fast or slow) is determined by the unit, on the basis of the remaining
battery charge and the ambient temperature).
Notes :
→
If the 5-position rotary switch is set to the "OFF" position while the battery is being charged, the
"Bat. chrg." message is displayed if the unit is performing a fast charge and the battery level symbol
flashes. The "Bat. Full" message is displayed once the battery is fully recharged.
→
If the unit is used while it is being charged, only the battery level symbol will flash while the battery
is charging.
1.4 ENDURANCE
The endurance depends on the measurements performed :
Measurements
Consumption
Number of possible
measurements
Instrument on standby
~0.7 W
None
Voltage measurement
~ 1.65 W
2 500
Continuity measurement
(1)
~ 6 W
1 500
Earth measurement
(2)
~ 3.5 W
1 000
Resistivity measurement
(3)
~ 3.2 W
1 200
(1)
R = 1
Ω
and R
H
= R
S
= 1k
Ω
(2)
R
E
= 10
Ω
and R
H
= R
S
= R
E
= R
ES
= 1k
Ω
(3)
R
S-ES
= 100
Ω
1.5 OPERATING CONDITIONS
Temperature
Service range : -10 to +55°C
Measurement range : 0 to +35°C
Storage and transport range (without batteries) : -40 to +70°C
Relative humidity
(without condensation)
Service conditions : 80% max.
Storage conditions : 90% max.
Tightness
IP53 per NF standard EN 60 529