Automatic reversal of polarity for a measurement at 200 mA.
TEST
If you selected the pulse mode, press the
TEST
button once and the measurement stops automatically when it is over.
If you selected the permanent mode, press the
TEST
button once to start the measurement and a second time to
stop it.
Measurement current.
R =
CONTINUITY
02
/
10
/
2014 10
:
47
2
.
00
Ω
- - .- Hz
0 . 8
3
Ω
I 2 0 7 . 4 m A
R + 0 . 5 9
Ω
R - 1 . 0 8
Ω
1 %
..\..
1 %
Measurement with a positive cur-
rent (R+).
Case where the measurement is
below the alarm threshold.
3.3.4. READING OF THE RESULT
In the case of a 200 mA current:
To activate the alarm.
Measurement at positive polarity only.
Measurement at negative polarity only.
Value of the alarm threshold.
Measurement with reversal of polarity.
Once the parameters have been defined, you can start the measurement.
Measurement with a negative cur-
rent (R-).
Compensation for the resistance of
the measurement leads is activated.
Permanent mode.
002.00
k
Ω
Ω
To deactivate the alarm.
To set the alarm threshold (see §3.14). The default threshold is 2
W
.
Measurement result:
(R+) + (R-)
2
Use the
key to see the rest of the
measurement display.