Channel
Dual channels, isolated between channels
Pulse Amplitude
0 ~ 80mA = 0 ~ 40 Volts, adjustable (at 500 ohm load)
Pulse Frequency (Hz)
2 ~ 150
Pulse Width (µs)
50 ~ 300
Waveform
Asymmetric biphasic square pulse.
Timer Control (mins)
5 ~ 90 mins or continuous
Stimulation Modes
Burst, Normal, MRW, SD, and Bi-Pulse
Power Supply
9V DC square shape battery
Size (D x W x H)
0.9" x 2.3" x 4.0" (24 mm x 59 mm x 101 mm)
Weight (including battery)
4.4 oz (124 g)
Safety standard
EN EN60601-1, EN EN60601-1-2, IEC 60601-2-10
Operation Ambient Temperature Range
50 ~ 95º F (10 ~ 35ºC)
Operation Ambient Humidity Range
20 ~ 90% RH
Storage & Transportation Temperature Range 32 ~ 158ºF (0 ~ 70ºC)
Storage & Transportation Humidity Range
20 ~ 90% RH
*All values have
10% tolerance.
22
4.15 Battery:
To replace the battery, remove the battery cover and extract the battery. Replace
it with a 9 V alkaline or similar rechargeable battery. Make sure the battery is
inserted correctly.
4.16 Care of Electrodes:
To avoid skin irritation and ensure good contact with skin, clean silicone rubber
electrodes with soap and water frequently. The electrodes must be dried
completely before using.
❈
If you are using self-adhesive electrodes, disregard this procedure.
❈
Always use the electrodes with the requirements of the EN60601-1 and EN60601-2, such as with CE
mark, or which are legally marketed in the US under 510(K) procedure.
4.17 Care of Electrode cords:
Clean the electrode cords by wiping them with damp cloth. Coating them lightly with talcum
powder will reduce tangles and prolong the life.
21
6. SPECIFICATION
5. HANDLING AND STORAGE
Keep this device in the carrying case and store at room temperature.