14
4.6 Select the mode:
Press "MODE" button to set the stimulation mode recommended by your
physician or therapist. For details about stimulating waveform and
sequences, please refer to Sec. 3 "Stimulation Modes".
Caution: Consult physicians for your suitable stimulation mode.
4.7 Adjust the Pulse Rate:
The pulses rate are adjustable 2~150Hz. 2Hz~20 Hz in 1 Hz increment
and 20~ 150 Hz in 5 Hz increment.
Press the SET button to enter the Pulse Rate set function, then press
INCREASE or DECREASE buttons to adjust Pulse Rate to the setting
recommended by your medical professional.
4.4 Place electrodes on skin:
Apply electrodes to the exact site indicated by your physician following the
instruction included with the electrodes labeling. Before applying
electrodes, be sure the skin surface over which electrodes are placed is
thoroughly cleaned and dried. Make sure the electrodes are placed firmly
to skin and make good contact between the skin and the electrodes. Place
the electrodes over the skin; attach them properly, firmly and evenly.
ADJUSTING THE CONTROLS
4.5 Adjust Output:
Turn Amplitude Control knob for Channel 1 or 2 clockwise, then you will
hear a "beep" sound. Before you increase the Amplitude, you must select
the mode, rate and width.
Caution: If the stimulation levels are uncomfortable or become uncomfortable, reduce
the stimulation intensity to a comfortable level and contact your medical
practitioner if problems persist.
13