58
Intelect® Mobile Ultrasound
MAINTENANCE
Maintaining the intelect Mobile UltrasoUnd
The following items should be checked at least monthly to ensure proper operation of this unit:
• Power cord and plug: Check to make sure the cord is not frayed, kinked, and does not have torn or cut insulation.
• Applicator cable: Check to make sure the cable is flexible, free of kinks, not frayed, and the insulation is intact.
• Sound head face: Check to see that there is no build-up of gel or foreign material on the aluminum face.
Cleaning
To clean the accessories, use only soap and water . Alcohol may be used to disinfect the aluminum surface, but avoid the plastic area .
The Intelect Mobile Ultrasound case may be cleaned by wiping with a damp cloth or mild cleaning solution . Avoid abrasive
cleansers .
NOTE: The sound head must be cleaned with alcohol between each therapy session .
Service
The Intelect Mobile Ultrasound must be recalibrated annually . It is recommended that all DJO ultrasound products be returned to
the factory or an authorized servicing dealer for repairs or recalibration . Recalibration is also recommended after the replacement or
repair of any major component . Should the unit require service, warranty, or repair, please contact the selling dealer or your local DJO
customer service .
EU Directive on Waste Electrical and Electronic Equipment (WEEE), ensures that product is appropriately disposed of or recycled at
the end of its life .