5 7
MAINTENANCE
E N
INTELECT
®
MOBILE 2 ULTRASOUND USER MANUAL
CALIBRATION REQUIREMENTS
The unit was calibrated during the manufacturing process
and doesn't need calibration during the product life.
DEVICE DISPOSAL
Council Directive 2012/19/EU concerning Waste Electrical
and Electronic Equipment (WEEE) requires not to dispose of
WEEE as municipal waste. Contact your local distributor for
information regarding disposal of the unit and accessories.
INSTRUCTION FOR SOFTWARE UPGRADE
1.
Go to the Chattanooga website
www.chattanoogarehab.com
2.
Go to Intelect Mobile 2 product tab
3.
Complete the registration form to be informed about
new product software version availability and IFU
updates (if not already done before)
4. Go to documents tab
5.
Download firmware upgrade onto USB stick
6. Switch OFF the device
7.
Insert USB key
8. Switch ON the device
9. Device will automatically detect firmware update
availability
10. Once firmware update is finished, Home screen will be
displayed. Device is ready for use.
IFU DOWNLOAD
1.
Go to the Chattanooga website
www.chattanoogarehab.com
2.
Go to Intelect Mobile 2 product tab
3.
Complete the registration form to be informed about
new product software version availability and IFU
updates if not already done before
4. Go to documents tab
5.
Click on the latest version of your Intelect Mobile 2
device (COMBO, US or STIM) User manual to download
Nota: a pdf viewer is required to display IFU
A hard copy of the IFU can be requested from DJO either by
registration on the website or you local DJO office or dealer,
the copy will be delivered to you within 7 days
INSTALLATION OF BATTERY
1.
Unscrew the battery door on the bottom of the device
(2 screws)
2.
Remove the battery door
3.
Plug the new battery to the battery connector
4. Insert the battery in its location
5.
Replace the battery door with the 2 screws
REPLACEMENT BATTERY
1.
Unscrew the battery door on the bottom of the device
(2 screws)
2.
Remove the battery door
3.
Unplug and remove the battery
4. Plug the new battery into the battery connector
5.
Insert the battery in its location
6. Replace the battery door with the 2 screws
Note: in case of unused device with the battery installed, it is
recommended to connect the device to the mains power and power on
the device with the main ON/OFF switch on the back of the device at
least once every 4 months to allow the battery to recharge.